<9. General Specifications>
9-7
IM 01C25F01-01E
Model
Suffix Codes
Description
EJA510E
EJA530E
. . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . .
Absolute pressure transmitter
Gauge pressure transmitter
Output
signal
-D . . . . . . . . . . . . . . . . . . . .
-J . . . . . . . . . . . . . . . . . . . .
-F . . . . . . . . . . . . . . . . . . . .
-G . . . . . . . . . . . . . . . . . . . .
-Q . . . . . . . . . . . . . . . . . . . .
4 to 20 mA DC Output with digital communication (BRAIN protocol)
4 to 20 mA DC Output with digital communication
(HART 5/HART 7 protocol)
*1
Digital communication (F
OUNDATION
Fieldbus protocol,
refer to IM 01C25T02-01EN)
Digital communication (PROFIBUS PA protocol,
refer to IM 01C25T04-01EN)
Low Power, 1 to 5 V DC with digital communication (HART 7 protocol)
Measurement
span (capsule)
A. . . . . . . . . . . . . . . . . . .
B. . . . . . . . . . . . . . . . . . .
C. . . . . . . . . . . . . . . . . . .
D. . . . . . . . . . . . . . . . . . .
10 to 200 kPa (1.45 to 29 psi)
0.1 to 2 MPa (14.5 to 290 psi)
0.5 to 10 MPa (72.5 to 1450 psi)
5 to 50 MPa (720 to 7200 psi)
*10
Wetted parts
material
*2
S. . . . . . . . . . . . . . . . . .
H. . . . . . . . . . . . . . . . . .
Process connector
Diaphragm
Others
316L SST
Hastelloy C-276
*3
316L SST
Hastelloy C-276
*3
Hastelloy C-276
*3
Hastelloy C-276
*3
Process
connections
4 . . . . . . . . . . . . . . . .
7 . . . . . . . . . . . . . . . .
8 . . . . . . . . . . . . . . . .
9 . . . . . . . . . . . . . . . .
P. . . . . . . . . . . . . . . .
1/2 NPT female
1/2 NPT male
G1/2 DIN 16 288 male
*4
M20×1.5 DIN 16 288 male
*4
For Diaphragm Seal System. Refer to IM 01C25W01-01EN.
—
N. . . . . . . . . . . . . .
Always N
—
-0 . . . . . . . . . . .
Always 0
Amplifier housing
►
1 . . . . . . . . . .
2 . . . . . . . . . .
3 . . . . . . . . . .
Cast aluminum alloy
Cast aluminum alloy with corrosion resistance properties *
5
ASTM CF-8M stainless steel
*6
Electrical connection
►
0 . . . . . . . . .
2 . . . . . . . . .
4 . . . . . . . . .
5 . . . . . . . . .
7 . . . . . . . . .
9 . . . . . . . . .
A. . . . . . . . .
C. . . . . . . . .
D. . . . . . . . .
G1/2 female, one electrical connection without blind plugs
1/2 NPT female, two electrical connections without blind plugs
M20 female, two electrical connections without blind plugs
G1/2 female, two electrical connections with a blind plug
*7
1/2 NPT female, two electrical connections with a blind plug
*7
M20 female, two electrical connections with a blind plug
*7
G1/2 female, two electrical connections and a 316 SST blind plug
1/2 NPT female, two electrical connections and a 316 SST blind plug
M20 female, two electrical connections and a 316 SST blind plug
Integral indicator
►
D. . . . . . .
E. . . . . . .
N. . . . . . .
Digital indicator
*9
Digital indicator with the range setting switch (push button)
*8
None
Mounting bracket
►
L . . . . .
N. . . . .
316 SST 2-inch pipe mounting
None
Optional codes
/
Optional specification
The “►” marks indicates the most typical selection for each specification. Example: EJA530E-DAS4N-012NN/
.
*1:
HART 5 or HART 7 is selectable. Specify upon ordering.
*2:
Users must consider the characteristics of selected wetted parts material and the influence of process fluids. The use of
inappropriate materials can result in the leakage of corrosive process fluids and cause injury to personnel and/or damage to plant
facilities. It is also possible that the diaphragm itself can be damaged and that material from the broken diaphragm and the fill fluid
can contaminate the user’fs process fluids.
Be very careful with highly corrosive process fluids such as hydrochloric acid, sulfuric acid, hydrogen sulfide, sodium hypochlorite,
and high-temperature steam (150°C [302°F] or above). Contact Yokogawa for detailed information of the wetted parts material.
*3:
Hastelloy C-276 or ASTM N10276.
*4:
Not applicable for combination of capsule code D and wetted parts material code H.
*5:
Not applicable for electrical connection code 0, 5, 7, 9 and A. Content rate of copper in the material is 0.03% or less and content rate
of iron is 0.15% or less.
*6:
Not applicable for electrical connection code 0, 5, 7 or 9.
*7:
Material of a blind plug is aluminum alloy or 304 SST.
*8:
Not applicable for output signal code F.
*9:
Not applicable for output signal code G.
*10: 5 to 70 MPa (720 to 10150 psi) when /HG is specified.