5-51
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:ANALysis:FFT<x>:MEASure:PEAK:
CALCulation:STATe<x>
Function Turns ON/OFF the calculation items of the FFT
computation or queries the current setting.
Syntax
:ANALysis:FFT<x>:MEASure:PEAK:
CALCulation:STATe<x> {<Boolean>}
:ANALysis:FFT<x>:MEASure:PEAK:
CALCulation:STATe<x>?
<x> of FFT<x> = 1 or 2
<x> of STATe<x> = 1 to 4
Example
:ANALYSIS:FFT1:MEASURE:PEAK:
CALCULATION:STATE1 ON
:ANALYSIS:FFT1:MEASURE:PEAK:
CALCULATION:STATE1? -> :ANALYSIS:FFT1:
MEASURE:PEAK:CALCULATION:STATE1 1
:ANALysis:FFT<x>:MEASure:PEAK:
CALCulation:VALue<x>?
Function Queries the measured value of the calculation item of
the FFT computation.
Syntax
:ANALysis:FFT<x>:MEASure:PEAK:
CALCulation:VALue<x>?
<x> of FFT<x> = 1 or 2
<x> of VALue<x> = 1 to 4
Example
:ANALYSIS:FFT1:MEASURE:PEAK:
CALCULATION:VALUE1? -> :ANALYSIS:FFT1:
MEASURE:PEAK:CALCULATION:
VALUE1 1.000E+00
:ANALysis:FFT<x>:RANGe
Function Sets the measurement source window used in the
FFT computation or queries the current setting.
Syntax
:ANALysis:FFT<x>:RANGe {MAIN|Z1|Z2}
:ANALysis:FFT<x>:RANGe?
<x> = 1 or 2
Example
:ANALYSIS:FFT1:RANGE MAIN
:ANALYSIS:FFT1:RANGE? -> :ANALYSIS:
FFT1:RANGE MAIN
:ANALysis:FFT<x>:RPARt (Real Part)
Function Sets the source trace of the real part of the FFT
computation or queries the current setting.
Syntax
:ANALysis:FFT<x>:RPARt {<NRf>}
:ANALysis:FFT<x>:RPARt?
<x> = 1 or 2
<NRf> = 1 to 8
Example
:ANALYSIS:FFT1:RPART 1
:ANALYSIS:FFT1:RPART? -> :ANALYSIS:
FFT1:RPART 1
:ANALysis:FFT<x>:RPOSition (Ref
Position)
Function Sets the center point of magnification of the vertical
axis of the FFT computation or queries the current
setting.
Syntax
:ANALysis:FFT<x>:RPOSition {<NRf>}
:ANALysis:FFT<x>:RPOSition?
<x> = 1 or 2
<NRf> = –4 to 4 (div)
Example
:ANALYSIS:FFT1:RPOSITION 1
:ANALYSIS:FFT1:RPOSITION? -> :ANALYSIS:
FFT1:RPOSITION 1.000E+00
:ANALysis:FFT<x>:VERTical?
Function Queries all settings related the vertical axis of the
FFT computation.
Syntax
:ANALysis:FFT<x>:VERTical?
<x> = 1 or 2
Example
:ANALYSIS:FFT1:VERTICAL? -> :ANALYSIS:
FFT1:VERTICAL:
LEVEL 1.000E+00;MODE AUTO;
SENSITIVITY 1.000E+00
:ANALysis:FFT<x>:VERTical:LEVel
Function Sets the display position of the vertical axis of the
FFT computation or queries the current setting.
Syntax
:ANALysis:FFT<x>:VERTical:LEVel {<NRf>}
:ANALysis:FFT<x>:VERTical:LEVel?
<x> = 1 or 2
<NRf> = –1.000E+31 to 1.000E+31 (dBV)
Example
:ANALYSIS:FFT1:VERTICAL:LEVEL 1
:ANALYSIS:FFT1:VERTICAL:LEVEL?
-> :ANALYSIS:FFT1:VERTICAL:
LEVEL 1.000E+00
:ANALysis:FFT<x>:VERTical:MODE
Function Sets the vertical axis mode of the FFT computation or
queries the current setting.
Syntax
:ANALysis:FFT<x>:VERTical:MODE {AUTO|
MANual}
:ANALysis:FFT<x>:VERTical:MODE?
<x> = 1 or 2
Example
:ANALYSIS:FFT1:VERTICAL:MODE AUTO
:ANALYSIS:FFT1:VERTICAL:MODE?
-> :ANALYSIS:FFT1:VERTICAL:MODE AUTO
5.3 ANALysis Group