5-94
IM 701310-17E
:GONogo:TELecomtest:SELect<x>:MASK:
ELEMent<x>:SPCount
(Sample Point Count)
Function Sets the upper and lower limits of the number of
sampled data points for the element that results in
error or queries the current setting.
Syntax
:GONogo:TELecomtest:SELect<x>:MASK:E
LEMent<x>:SPCount {<NRf>,<NRf>}
:GONogo:TELecomtest:SELect<x>:MASK:
ELEMent<x>:SPCount?
<x> of SELect<x> = 1 to 4
<x> of ELEMent<x> = 1 to 4
<NRf> = See the DL9000 User’s Manual.
Example
:GONOGO:TELECOMTEST:SELECT1:MASK:
ELEMENT1:SPCOUNT 1,2
:GONOGO:TELECOMTEST:SELECT1:MASK:
ELEMENT1:SPCOUNT? -> :GONOGO:
TELECOMTEST:SELECT1:MASK:
ELEMENT1:SPCOUNT 2.000E+00,1.000E+00
:GONogo:TELecomtest:SELect<x>:MASK:
ELEMent<x>:WCOunt (Wave Count)
Function Sets the upper and lower limits of the number of
acquisitions for the element that results in error or
queries the current setting.
Syntax
:GONogo:TELecomtest:SELect<x>:MASK:
ELEMent<x>:WCOunt {<NRf>,<NRf>}
:GONogo:TELecomtest:SELect<x>:MASK:
ELEMent<x>:WCOunt?
<x> of SELect<x> = 1 to 4
<x> of ELEMent<x> = 1 to 4
<NRf> = See the DL9000 User’s Manual.
Example
:GONOGO:TELECOMTEST:SELECT1:MASK:
ELEMENT1:WCOUNT 1,2
:GONOGO:TELECOMTEST:SELECT1:MASK:
ELEMENT1:WCOUNT? -> :GONOGO:TELECOMTEST:
SELECT1:MASK:
ELEMENT1:WCOUNT 2.000E+00,1.000E+00
:GONogo:ZPARameter?
Function Queries all settings related to zone/parameter
determination.
Syntax
:GONogo:ZPARameter?
Example
:GONOGO:ZPARAMETER? -> :GONOGO:
ZPARAMETER:SELECT1:
MODE PARAMETER;PARAMETER:CATEGORY FFT;
FFT1:CALCULATION1 0.000E+00,1.000E+00;:
GONOGO:ZPARAMETER:SELECT1:RECTANGLE:
HORIZONTAL 0.000E+00,1.000E+00;
VERTICAL 0.000E+00,1.000E+00;:GONOGO:
ZPARAMETER:SELECT1:TRACE 1;WAVE:
TRANGE 1.000E+00,2.000E+00;:GONOGO:
ZPARAMETER:SELECT1:WINDOW MAIN;:GONOGO:
ZPARAMETER:SELECT2:MODE PARAMETER;
PARAMETER:CATEGORY FFT;;FFT1:
CALCULATION1 0.000E+00,1.000E+00;:
GONOGO:ZPARAMETER:SELECT2:RECTANGLE:
HORIZONTAL 0.000E+00,1.000E+00;
VERTICAL 0.000E+00,1.000E+00;:GONOGO:
ZPARAMETER:SELECT2:TRACE 1;WAVE:
TRANGE 1.000E+00,2.000E+00;:GONOGO:
ZPARAMETER:SELECT2:WINDOW MAIN;:GONOGO:
ZPARAMETER:SELECT3:MODE PARAMETER;
PARAMETER:CATEGORY FFT;FFT1:
CALCULATION1 0.000E+00,1.000E+00;:.....
:GONogo:ZPARameter:SELect<x>?
Function Queries all settings related to the condition of the
zone/parameter determination.
Syntax
:GONogo:ZPARameter:SELect<x>?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:SELECT1? -> :GONOGO:
ZPARAMETER:SELECT1:
MODE PARAMETER;PARAMETER:CATEGORY FFT;
FFT1:CALCULATION1 0.000E+00,1.000E+00;:
GONOGO:ZPARAMETER:SELECT1:RECTANGLE:
HORIZONTAL 0.000E+00,1.000E+00;
VERTICAL 0.000E+00,1.000E+00;:GONOGO:
ZPARAMETER:SELECT1:TRACE 1;WAVE:
TRANGE 1.000E+00,2.000E+00;:GONOGO:
ZPARAMETER:SELECT1:WINDOW MAIN
:GONogo:ZPARameter:SELect<x>:MODE
Function Sets the mode of the condition or queries the current
setting.
Syntax
:GONogo:ZPARameter:SELect<x>:MODE {PARa
meter|POLYgon|RECTangle|WAVE}
:GONogo:ZPARameter:SELect<x>:MODE?
<x> = 1 to 4
Example
:GONOGO:ZPARAMETER:SELECT1:MODE WAVE
:GONOGO:ZPARAMETER:SELECT1:MODE?
-> :GONOGO:ZPARAMETER:SELECT1:MODE WAVE
5.12 GONogo Group