2-23
IM 701730-51E
SPI Bus Signal Analysis Function
2
Note
The detailed analysis list can be saved directly to an external storage medium in ASCII format
(.txt extension). For details, see section 2.6, “Saving the Data of the Detailed Analysis List.”
Setting the Search Condition
29.
Press the Data Search soft key. The Data Search menu appears.
30.
Press the Search Setup soft key. The Search Setup dialog box opens.
31.
Use jog shuttle & SELECT to set the search type to Frame Pattern or Indefinite
State (Type box).
If you select Indefinite State, proceed to step 33.
32.
Use jog shuttle & SELECT to set the notation system used to set the
determination pattern to Hex or Bin (Pattern Format box). The format in the
Data Pattern box changes accordingly.
33.
Use jog shuttle & SELECT to set the data to be searched to Data1 or Data2
(Source box).
If you selected Indefinite State in step 31, proceed to step 36.
34.
Use jog shuttle & SELECT to set the data length of the determination pattern to
be specified from 1 to 8 bytes (Data Byte box). The number of bytes displayed
in the Data Pattern box is set to the number of bytes that matches the selected
result.
35.
Use jog shuttle & SELECT to set the determination pattern for each byte in
hexadecimal or binary (Data Pattern box). When determination is not to be
performed, select X.
36.
Press ESC. The Detail dialog box closes.
Set the notation system of the data to be searched
to Hex or Bin. (The format in the Data Pattern box is
set to the format that matches the selected result.)
Set the search type to Frame Pattern or Indefinite State.
(If Indefinite State is selected, select the search target data in the Source box.)
Set the determination pattern per byte using hexadecimal or binary values.
Set the data to be searched to Data1 or Data2.
Select the search unit (data length) in the range of
1 to 8 bytes. (The number of bytes displayed in the
Data Pattern box is set to the number of bytes that
matches the selected result.)
2.5 Analyzing/Searching Data