1-26
IM 701730-51E
• Setting the Data Channel
8.
Use job shuttle & SELECT to set the data channel to CH2 or CH4 (Data CH
box).
The clock channel and data channel settings are synchronized. If you set the clock
channel to CH1, the data channel is set to CH2; if you set the clock channel to CH3,
the data channel is set to CH4.
Setting the Threshold Level
9.
Use jog shuttle & SELECT to set the level used to determine a high level signal
(Thr Upper box).
10.
Likewise, use jog shuttle & SELECT to set the level used to determine a low
level signal (Thr Lower box).
Set the data channel to CH2 or CH4.
Set the level used to determine high/low.
• Setting the Analysis Reference Point
11.
Use jog shuttle & SELECT to select the method of setting the analysis
reference point: Trigger Position or Manual.
Highlighting of the mark to the left of each item indicates that it is used as an analysis
reference point.
When Manually Setting the Analysis Reference Point
12.
Use jog shuttle & SELECT to set the analysis reference point in the range of
–5 to +5 [div] (Manual box).
When setting the analysis reference point manually, you can enable the translucent
display. This allows you to set the analysis reference point while viewing the displayed
waveform. For details on the translucent display, see section 8.10, “Turning ON/OFF
the Translucent Display” in the
DL1700E Series User’s Manual (IM701730-01E).
Select this to make the trigger position
the analysis reference point.
To manually set the analysis reference point, select
this and set the value in the range of –5 to +5 [div].
13.
Press ESC. The Analyze Setup dialog box closes.
1.5 Analyzing/Searching Data