1-8
IM 701730-51E
• When I
2
C -> Pattern Is Selected
24.
Press the Set Pattern soft key. The Set Pattern dialog box displays the
assignment conditions of the I
2
C Bus signal of CH1 and CH2 and the setup
screen for the patterns of CH3 to CH4.
25.
Use jog shuttle & SELECT to set the clock channel to None or any of the
channels from CH3 to CH4 (Clock CH box).
26.
Use jog shuttle & SELECT to set the patterns of CH3 to CH4 using H (high), L
(low), and X. For the channel set as the clock channel in step 25, set the slope
to (rising edge) or (falling edge).
27.
If you set the clock channel to None in step 25, use jog shuttle & SELECT to
set the pattern condition to Enter or Exit (Condition box).
Select the clock channel
from None.
When Clock CH is set to None
When Clock CH is set to any channel
from CH3 to CH4
* Set the slope of the channel selected
for the clock channel to or .
Select the CH3 to CH4
pattern* using H, L, and X.
Set the pattern condition
to Enter or Exit.
Select the clock
channel from CH3 to
CH4.
Select the CH3 to CH4
pattern* using H, L, and
X.
28.
Press ESC. The Set Pattern dialog box closes.
Proceed to step 29.
Note
If you set Combination to I
2
C -> Pattern and set the patterns of CH3 to CH4 to all Xs, triggers
will not be activated. If the clock channel is set to any of the channels from CH3 to CH4, the
pattern conditions of other channels are always set to true.
1.4 Setting the I