![YOKOGAWA CA450 User Manual Download Page 69](http://html.mh-extra.com/html/yokogawa/ca450/ca450_user-manual_911486069.webp)
67
RA
Sets or queries the current sweep output
Normal Cal.
Command = RAm<CRLF> -> Answer = RAm<CRLF>
Command = RA?<CRLF> -> Return = RAm<CRLF>
Parameters
m = 0: Slow Linear (default)
1: Fast Linear
2: Slow Step
3: Fast Step
When the CA450 is in a measurement mode or constant
current output mode, ERR13 is returned.
RG
Sets or queries the auto range and range hold settings
Normal Cal.
Command = RGm<CRLF> -> Answer = RGm<CRLF>
Command = RG?<CRLF> -> Return = RGm<CRLF>
Parameters
m = 0: Auto range (default)
1: Range hold
When the CA450 is in an output mode or DCmV, ACmV, diode
test, or continuity check mode, ERR13 is returned.
Î
RP
Sets or queries the RELΔ/% display
Normal Cal.
Command = RPm<CRLF> -> Answer = RPm<CRLF>
Command = RP?<CRLF> -> Return = RPm<CRLF>
Parameters
m = 0: Normal measurement (default)
1: RELΔ display
2: REL% display
If the CA450 is in an output mode, ERR13 is returned.
Î