6
MR
Sets or queries the measurement range
Normal Cal.
Command = MRm<CRLF> -> Answer = MRm<CRLF>
Command = MR?<CRLF> -> Return = MRm<CRLF>
Parameters
[DCV] m =
1: 6 V
3: 60 V
4: 600 V
5: 1000 V
[DCA] m =
0: 30 mA
1: 100 mA
[ACV] m =
0: 6 V
1: 60 V
2: 600 V
3: 1000 V
[Ω] m =
0: 600 Ω
1: 6 kΩ
2: 60 kΩ
3: 600 kΩ
4: 6 MΩ
5: 60 MΩ
[DCmV] m =
0: 600 mV
[ACmV] m =
0: 600 mV
[Diode test] m =
0: 2 V
[Continuity check] m = 0: 600 Ω
[LOOP POWER] m = 0: 30 mA