6
CMF
Queries the measurement mode being calibrated
Normal Cal.
Command = CMF?<CRLF> -> Return = CMFm<CRLF>
Parameters
m = 0: DCV
2: Ω
5: Hz
7: ACV
8: DCmV
9: ACmV
10: Diode test
11: Continuity check
12: DCmA
13: LOOP POWER
If the CA450 is in constant current output, current sweep
output, or continuity check function, ERR13 is returned.
Î
CP
Sets the calibration point
Normal Cal.
Command = CPm<CRLF> -> Answer = CPm<CRLF>
Parameters
m = 0: +FS calibration
1: Peak hold zero calibration or +5% output
calibration
2: –FS calibration
3: –5% calibration
–FS calibration and –5% calibration are used to calibrate the
SIMULATE output mode.
If –FS calibration or –5% calibration is specified in
measurement mode, ERR13 is returned.