BZ
Sets or queries whether the buzzer is on or off
Normal Cal.
Command = BZm<CRLF> -> Answer = BZm<CRLF>
Command = BZ?<CRLF> -> Return = BZm<CRLF>
Parameters
m = 0: OFF
1: On (default)
Î
CC
Sets or queries whether the measurement mode is DCmV or
ACmV
Normal Cal.
Command = CCm<CRLF> -> Answer = CCm<CRLF>
Command = CC?<CRLF> -> Return = CCm<CRLF>
Parameters
m = 0: DCmV measurement (default)
1: ACmV measurement
In SENSOR mode
m = 0: DC sensor measurement (default)
1: AC sensor measurement
If the measurement mode is neither DCmV nor ACmV, ERR13
is returned.
CD
Confirms the output calibration value
Normal Cal.
Command = CD<CRLF> -> Answer = CD<CRLF>
In output calibration mode, the current output settings are
used as the output calibration values for the currently selected
function, range, and calibration point (+FS, +5%, -FS, or -5%).
If the CA450 is in a measurement mode, ERR13 is returned.
Î
CL
Queries the calibration item
Normal Cal.
Command = CL?<CRLF> -> Return = CLm<CRLF>
Parameters
m = 3: Output calibration
4: Measurement calibration
Î