7
CR
Sets the output calibration value
Normal Cal.
Command = CRm<CRLF> -> Answer = CRm<CRLF>
Command = CR?<CRLF> -> Return = CRm<CRLF>
Parameters
m = 18.0000 to 22.0000 for ±FS calibration
m = 0.9600 to 1.0400 for ±5% calibration
Î
CS
Confirms the measurement calibration value
Normal Cal.
Command = CS<CRLF> -> Answer = CS<CRLF>
In measurement calibration mode, the current measured values
are used as the measurement calibration values for the currently
selected function, range, and calibration point (+FS or +0).
If the CA450 is in an output mode, ERR13 is returned.
Î
CSF
Queries the output mode being calibrated
Normal Cal.
Command = CSF?<CRLF> -> Return = CSFm<CRLF>
Parameters
m = 14: Constant current output
15: Current sweep output
If the CA450 is in a measurement mode, ERR13 is returned.
Î
CW
Saves the calibration value
Normal Cal.
Command = CW<CRLF> -> Answer = CW,OK<CRLF>
(normal completion)
Saves the data after a mode or range is calibrated. If you turn
off the CA450 without executing this command, the calibrations
made immediately before are discarded.
Î