24
IM AQ6373B-02EN
9 Specifications
For the specifications of the limited model (SUFFIX -02, -01), see IM AQ6373B-51EN, a manual specifically for the limited model
(SUFFIX -02, -01).
Item
Specifications
Applicable fiber
SM, GI (50/125 μm, 62.5/125 μm), Large core size fibers (core diameter of up to 800 µm)
Measurement wavelength range
1
350 to 1200 nm
Span
1
0.5 nm to 850 nm (entire wavelength range), 0 nm
Wavelength accuracy
1, 2
±0.2 nm (400 to 1100 nm)
±0.05 nm (633 nm)
Wavelength resolution setting
1, 3
0.02, 0.05, 0.1, 0.2, 0.5, 1.0, 2.0, 5.0, 10.0 nm, and 0.01 nm (at a measured wavelength
range of 400 to 470 nm)
Measurement data point
101 to 50001, AUTO
(Wavelength sampling points)
Wavelength sampling resolution 0.001 nm minimum
Level sensitivity setting
NORM_HOLD, NORM_AUTO, NORMAL, MID, HIGH1, HIGH2 and HIGH3
High dynamic range mode SWITCH (Sensitivity: MID, HIGH1, HIGH2, HIGH3)
Level sensitivity
4, 5
-80 dBm (500 to 1000 nm, resolution: 0.2 nm or more, measuring sensitivity: HIGH3)
-60 dBm (400 to 500 nm, resolution: 0.2 nm or more, measurement sensitivity: HIGH3)
-60 dBm (1000 to 1100 nm, resolution: 0.2 nm or more, measurement sensitivity:HIGH3)
Level accuracy
4
±1.0 dB (850 nm, -20 dBm, resolution: 0.2 nm or more, sensitivity setting: MID, HIGH1,
HIGH2, HIGH3) (when using an (MFD 5 µm @850 nm, NA 0.14 (typ.) SMF)
Leval linearity
4
±0.2 dB (input level: -40 to +0 dBm, measuring sensitivity: HIGH1, HIGH2, HIGH3)
Safe max. input power
4
+20 dBm (total light input power at 550 to 1100 nm)
+10 dBm (total light input power at 400 to 550 nm)
Dynamic range
1
60 dB or more (633 nm, ±0.5 nm of peak wavelength, resolution: 0.02 nm)
Sweep time
1, 6
0.5 s (NORM_AUTO), 1 s (NORMAL)
2
s
(MID)
5 s (HIGH1), 20 s (HIGH2), 75 s (HIGH3)
Function
Measurement
Setting of measuring conditions Center wavelength, span, wavelength sampling points, wavelength resolution, measurement
sensitivity, high dynamic mode, average count (1 to 999), double-speed measurement mode,
smoothing, Large core size fiber mode
Sweep settings
Single sweep, repeat sweep, AUTO (automatically sets measuring conditions), sweep
between marker, data logging
Measurement function
CW measurement, pulse light measurement, external trigger measurement, gate
measurement, air/vacuum wavelength measurement
Other
Sweep status output, analog output
Display
Vertical scale
Level scale (0.1 to 10 dB/div, linear), level subscale (0.1 to 10 dB/div, linear), reference level
display, DIV display (8, 10, 12), percentage display, dB/km display, power density (dB/nm)
display, noise mask
Horizontal scale
Horizontal wavelength (nm) display, frequency (THz) display, zoom in/zoom out display
Display mode and items Single waveform display, split screen display, data table display, label display, template
display, measurement condition display
Trace
Display function
Simultaneous display of 7 independent traces, max/min value detection display, calculation
between traces display, normalized display, roll averaging (sweep average) display (2 to 100
times), curve fit display, peak curve fit display, marker curve fit display
Other
Trace copy, trace clear, write mode fixed mode setting, show/hide setting
Marker and search
Marker
Delta marker (1024 points max.), vertical/horizontal line markers, advanced marker
Search
Peak search, bottom search, auto search (ON/OFF), search between vertical axis line
markers, search within zoom area