3-9
IM 2560A-01EN
Preparation
3
2
1
4
5
6
7
8
9
10
11
12
13
14
15
16
App
Index
Operations Performed When the Power Is Turned Off
Setup parameters immediately before (approx. 1 s before) the power switch is turned off is stored. The
same is true when the power cord is disconnected from the outlet. The next time the power is turned
on, the instrument powers up using the stored settings (for the settings that are stored, see appendix 7).
CAUTION
Do not turn off the power when the instrument’s output is turned on. Doing so can damage the
2560A. It can also cause damage to the devices connected to the 2560A. Turn the output off
first, and the turn off the power.
French
ATTENTION
Ne pas tourner l’appareil hors tension quand la sortie de l’instrument est activée. Cela peut
endommager le 2560A. Il peut aussi endommager les appareils connectés à la 2560A.
Tourner la sortie première et la coupez l’alimentation.
3.4 Turning On and Off the Power Switch