7-2
IM 2560A-01EN
7.2 Deviation Output
Procedure
Setting the Deviation
Turn the
deviation
dials 1 and 2 to set the deviation.
• The deviation display shows the specified deviation.
• The output display shows the output value with the deviation applied.
• Turn deviation dials 1 and 2 to the right to decrease the deviation and increase the output. Turn them to
the left to increase the deviation and decrease the output.
Deviation dial 2
Deviation dial 1
Set the second decimal digit at a resolution of 0.01.
Set the first decimal digit at a resolution of 0.2.
For both dials, carrying over and borrowing occur automatically
within the selectable range.
Setting a Preset
Flip the
PRESET
switch up or down to select a preset.
The selected PRESET indicator turns on, and the deviation display shows the preset deviation.
PRESET switch
Select a preset.
Presets
Explanation
Deviation output is a feature that adds or subtracts deviation from the voltage, current, resistance,
or temperature set with the main setting dials and outputs the result. Deviation and preset cannot be
used while sweeping or in high resolution mode.
Deviation
Set the deviation in the following range.
Selectable range: –20.00% to +20.00%, resolution: 0.01%
However, deviation may not exceed the source range.
• If the output is turned on, the output remains on even if you change the deviation with the deviation
dials.
• How the deviation changes when the main setting or the denominator m or numerator n of the
output divider is changed depends on the preset setting (explained later). See the description of
presets.