6-8
IM 2560A-01EN
Setting the Output Value Display Mode
1.
Turn the
RANGE
dial to select SETUP.
The SETUP menu appears.
RANGE dial
1. Select SETUP.
The SETUP indicator turns on.
2.
Turn a
main setting
dial to select TEMP.
“NEXT” appears on the deviation display. All the PRESET indicators turn on.
3.
Flip the
PRESET
switch down.
The TEMP menu appears. All the PRESET indicators turn off.
PRESET switch
3. Display the TEMP menu.
Main setting dials (use any)
2. Select TEMP.
PRESET indicator
4.
Turn a
main setting
dial to select MONI.
The deviation display shows the current output value display mode.
5.
Turn a
deviation dial
to select DEG or V.OHM.
• When you change the setting, all PRESET indicators turn on.
• If you decide not to change the setting, step 6 is not necessary.
6.
Flip the
PRESET
switch down.
The setting is confirmed. All the PRESET indicators turn off.
PRESET switch
6. Confirm the settings.
Main setting dials (use any)
4. Select MONI.
Deviation dials (either)
5. Select DEG or V.OHM.
PRESET indicator
6.4 Turning the Output On and Off and Setting the Output Value Display Mode