5-2
IM 2553A-01EN
5.2 Specifying the Output Setting and Polarity
Procedure
Specifying the Output Setting
Turn the
output setting
dials to set the source voltage, current, or resistance.
The display shows the specified voltage, current, or resistance.
Setting the Polarity
Flip the
POLARITY
switch up or down to set the polarity.
Example of setting 1.5000 V
Unit
Use each output setting dial to set the corresponding digit value,
shown by the frames.
Carrying over and borrowing occurs automatically within the
selectable range.
POLARITY switch
Polarity
Explanation
Selectable Range
The selectable range and resolution of each range are as follows. For detailed specifications, see
section 14.1.
As long as you remain in the same range, you can change the output setting and polarity even while
the output is on.
Voltage Ranges
Range
Source Range
Resolution
10 mV
±12.0000 mV
100 nV
100 mV
±120.000 mV
1 µV
1 V
±1.20000 V
10 µV
10 V
±12.0000 V
100 µV
30 V
± 32.000 V
1 mV
Current Ranges
Range
Source Range
Resolution
1 mA
±1.20000 mA
10 nA
10 mA
±12.0000 mA
100 nA
30 mA
± 32.000 mA
1 µA
100 mA
±120.000 mA
1 µA
Resistance Ranges
Range
Selectable Range
Resolution
400 Ω
18.00 Ω to 400.00 Ω
0.01 Ω