13-7
IM 2553A-01EN
Troubleshooting, Maintenance, and Inspection
3
2
1
4
5
6
7
8
9
10
11
12
13
14
App
Index
Other Error (350)
Code
Message
Corrective Action
Page
350
Queue overflow.
Read the error queue.
12-6
Note
Code 350 occurs when the error queue overflows.
Execution Errors (800 to 899)
Code Message
Corrective Action
Page
802
Invalid command.
CAL on the SETUP menu cannot be executed during
adjustment mode.
4-2
804
Invalid command.
Cannot be executed because the communication
command type is set to COMP (compatible mode).
7-7
805
Invalid command.
Communication commands cannot be executed because
the RANGE dial is set to SETUP.
—
System Errors (900/999)
Code
Message
Corrective Action
Page
911 to
999
CPU board error.
For details about the message and corrective action,
see section 13.3.
13-4
13.4 Communication Error Messages