Using the Auxiliary Inputs and Outputs
44
TUNICS T100S-HP User Manual
•
Make sure to use a 50
Ω
BNC cord to connect the TUNICS T100S-HP and the
waveform generator.
Figure 8: Low frequency modulation setup
2.
Set the operating current
I
bias
of the TUNICS T100S-HP in constant-current mode.
The relationship between the diode output current
I
diode
and the input voltage
V
in
is
then given by:
Formula:
Where:
•
V
in
is the input voltage applied to the LF Mod. BNC input connector (
V
in
must
always be positive and less than 5 V)
•
I
diode
is the total diode current
•
I
bias
is the polarization current of your TUNICS T100S-HP
•
R
is the 50
Ω
resistor of the LF modulation circuit
With a maximum allowed voltage input of 5 volts, the maximum input current you
can add to the TUNICS T100S-HP laser diode is 100 mA.
Resulting LF-
Modulated Diode
Current
The laser diode of your TUNICS T100S-HP is characterized by its saturation current (
I
max
)
and the cutoff current (
I
cutoff
).
To perform a diode intensity modulation under optimum conditions, you must ensure at
all times that the resulting modulated diode current obeys the following condition:
I
cutoff
≤
I
diode
≤
I
max
The
following
figure shows a behavioral representation the resulting modulated diode
current.
TUNICS T100S-HP Laser Diode Source
LF Modulation
Circuit
Waveform Generator
I
bias
Polarization Current
I
in
I
diode
V
in
BNC Cord
50
Ω
LF Mod.
BNC Connector
50
Ω
Idiode Ibias
2
Vin
R
--------
–
=
Summary of Contents for TUNICS T100S-HP
Page 2: ......
Page 6: ...About This Manual 6 TUNICS T100S HP User Manual...
Page 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...
Page 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...
Page 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...
Page 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...