Enabling Active Wavelength Monitoring
TUNICS T100S-HP User Manual
41
7.2
Performing Wavelength Monitoring
Wavelength
Monitoring
Scheme
The wavelength monitoring scheme consists in a loop-back to the tunable laser of the
wavemeter actual reading, and self-alignment of the tunable laser by iteration process to
achieve the user-defined target wavelength within the wavemeter’s accuracy. In order to
achieve the user-defined target wavelength within the wavemeter accuracy, the following
wavelength monitoring scheme is implemented:
•
The wavemeter actual reading loops back to the tunable laser
•
An iteration process self-aligns the tunable laser.
Figure 7: Wavelength Monitoring Scheme
Depending on the performance of the wavemeter you are using, the average iteration
time to reach the target wavelength may range from 1.5 s to 2 s.
The following procedure explains how to perform wavelength monitoring of the
TUNICS T100S-HP using a wavemeter.
Procedure
1.
Configure the wavemeter remote interface with the following parameters:
•
Remote: [On]
•
RS-232: [On]
•
Baud: [9600]
•
Handshaking: [None]
(TUNICS T100S-HP output)
Target wavelength:
λ
t
λ
1
out
TUNICS T100S-HP
Tunable External Cavity Laser
∆λ
n
λ
1
read
λ
n
out
∆λ
1
Wavemeter
λ
n
read
(User-setting)
Target wavelength:
λ
t
Summary of Contents for TUNICS T100S-HP
Page 2: ......
Page 6: ...About This Manual 6 TUNICS T100S HP User Manual...
Page 24: ...Installing and Connecting the Product 24 TUNICS T100S HP User Manual...
Page 36: ...Changing the Optical Output Power and Diode Current 36 TUNICS T100S HP User Manual...
Page 50: ...Using the Auxiliary Inputs and Outputs 50 TUNICS T100S HP User Manual...
Page 54: ...Verifying Performances 54 TUNICS T100S HP User Manual...
Page 60: ...Certification and Compliance 60 TUNICS T100S HP User Manual...