13
3. REMOTE OPERATION VIA COMMUNICATION
3.1: Start remote mode
・
Remote mode can start only when the controller is OFF state.
・
Start remote mode by specifying operation mode (fixed temp operation, program operation, auto
start, auto stop, or program auto start). However, when program operation or program auto start
is selected, remote mode cannot be started unless the following condition is satisfied.
Operation mode
Remote mode start conditions
Program operation Program patterns for the operation are specified.
Program auto start Program pattern for auto start is specified.
3.2: End remote mode
・
When remote mode ends, regardless of the remote operation state (standby/running/end) main
unit terminates operation and set remote operation state in standby, then quit remote mode.
After remote mode ends, the controller will be turned OFF (shows power OFF screen).
3.3: Remote operation start/stop
・
When starting remote operation during standby or at the end of operation, main unit starts
remote operation with operation mode and setting value set at that point.
・
Remote operation is stopped by transmitting remote operation stop command during remote
operation.
3.4: Operation mode change in remote mode
・
The operation mode can be changed to another during standby or at the end of operation.
However, when changing the operation mode to program operation or program auto-start, the
mode change cannot be completed unless the following condition is satisfied.
Operation mode
Operating mode change conditions
Program operation Program patterns for the operation are specified.
Program auto start Program pattern for auto start is specified.