1
1. COMMUNICATION SETTING PROCEDURES
1.1: Communication setting
Make appropriate settings to master device (workstation PC etc.) and to TYPE V controller on
main unit (hereafter, “main unit”)
The following table shows the communication setting items for TYPE V controller.
Set values on master device to match items 1-6 below on main unit
The following settings can be made by key operation.
Item
Item display
name
Communication settings
1
Communication
protocol
PRTCL
MODBUS-ASCII/RTU
2
Data length
DATA
7/8 bit
3
Parity
PARI
Non/odd/evn
4
Stop bit length
STOPB
1/2 bit
5
Communication rate
BPS
9600/19200/38400bps
6
Response delay time
DELAY
0, 10, 20-250 msec
7
Communication
address
ADDRS
1-99
8
Communication mode
CMODE
RW/R
: Factory defaults
Communication settings by key operation
1
Controller power should be turned off, showing only
the time.
If control panel shows anything other than time in the
sub (lower) display, press & hold
to turn off
the controller
So it only shows the time.
2
Press & hold
until "UPASS" shows in the sub
display and "00" shows in the main (upper) display.
Enter "11" using
and press
.
* Password is fixed at 11 and cannot be changed.