23
4. COMMUNICATION REFERENCE TABLE
Reference
No.
FNC
Code
R/W
Data name
Setting range
(in-transmission)
30108
04
R
Error status
BIT0 = temp sensor error
BIT1 = triac (SSR) short circuit
BIT2 = Heater disconnection
BIT3 = Fan error
BIT4 = Automatic damper error*
BIT5 = Independent overheat prevention device error
BIT6 = Three-phase power supply error*
BIT7 = Main relay contact failure
BIT8 = Refrigerator error*
BIT9 = RAM error
BIT10 = EEPROM error
BIT11 = Auto tune error*
BIT12 = External Error*
BIT13 = Inverter error*
BIT14 = High pressure error*
BIT15 = Overload error*
◆
Refer to main unit instruction manual for above
errors.
30109
04
R
Alert status
BIT0 = Water level error*
BIT1 = Door open*
BIT2 = Software overheat error
30110
04
R
Running
pattern/step No.
High order byte
Pattern No.
0-99 (0 = standby)
Low order byte
Step No.
0-99 (0 = standby)
1
3
5
4
2
0
7
6
0 = Alert OFF
1 = Alert ON
1
3
5
4
2
0
7
6
0 = No error
1 = Error occurs
9
11
13
12
10
8
15
14
Low order
High order