Cautions in Using with Safety 

Fundamental Matters of "WARNING!" and "CAUTION!" 

 

  WARNING! 

 

 

Do not use this unit in an area where there is flammable or explosive gas 

Never use this unit in an area where there is flammable or explosive gas. This unit is not explosion-
proof.    An  arc  may  be generated  when  the  power  switch  is  turned on  or  off,  and fire/explosion may 
result. (Refer to page 40 "List of Dangerous Substances".) 

 

 

Always ground this unit

 

Always ground this unit on the power equipment side in order to avoid electrical shock due to a power 
surge. 

 

 

If a problem occurs

 

If smoke or strange odor should come out of this unit for some reason, turn off the power key right away, 
and then turn off the circuit breaker and the main power.    Immediately contact a service technician for 
inspection. If this procedure is not followed, fire or electrical shock may result. Never perform repair work 
yourself, since it is dangerous and not recommended. 

 

 

Do not use the power cord if it is bundled or tangled

 

Do not use the power cord if it is bundled or tangled.    If it is used in this manner, it can overheat and 
fire may be caused. 

 

 

Do not process, bend, wring, or stretch the power cord forcibly

 

Do not process, bend, wring, or stretch the power cord forcibly.    Fire or electrical shock may result. 

 

 

Substances that can not be used

 

Never  use  explosive  substances,  flammable  substances  and  substances  that  include  explosive  or 
flammable ingredients in this unit. Explosion or fire may occur. (Refer to page 40 "List of Dangerous 
Substances"
.) 

 

  Do not disassemble or modify this unit

 

Do not disassemble or modify this unit.    Fire or electrical shock or failure may be caused. 

 

  Do not touch high-temperature parts 

The inside of the body or the door may become hot during and just after operation. It may cause burns. 

 

  CAUTION! 

 

 

During a thunder storm

 

During a thunderstorm, turn off

 

the power key immediately, then turn off the circuit breaker and the main 

power.    If this procedure is not followed, fire or electrical shock may be caused. 

 

 

Summary of Contents for DR 200

Page 1: ...his unit Keep this instruction manual around this unit for referring at anytime WARNING Carefully read and thoroughly understand the important warning items described in this manual before using this unit High Temperature Oven Model DR 200 Instruction Manual First Edition Yamato Scientific America Inc Santa Clara CA This paper has been printed on recycled paper ...

Page 2: ......

Page 3: ...haracters 14 Setting of Overheating Prevention Device 15 Fixed Temperature Operation 16 Quick Auto Stop Operation 17 Auto Stop Operation 18 Auto Start Operation 20 Program Operation 22 Other Functions 29 Handling Precautions 30 Maintenance Method 32 Daily Inspection and Maintenance 32 Long storage and disposal 33 When not using this unit for long term When disposing 33 In the Event of Failure 34 S...

Page 4: ...f a problem that may cause a serious accident or ev en f atality CAUTION If the caution is ignored there is the danger of a problem that may cause injury damage to property or the unit itself Meaning of Symbols This sy mbol indicates items that urge the warning including the caution A detailed warning message is shown adjacent to the sy mbol This sy mbol indicates items that are strictly prohibite...

Page 5: ...aution electrical shock Caution scald Caution no road heating Caution not to drench Caution water only Caution deadly poison Prohibit Prohibit generally Prohibit inflammable Prohibit to disassemble Prohibit to touch Compulsion Compulsion generally Compulsion connect to the grounding terminal Compulsion install on a flat surface Compulsion disconnect the power plug Compulsion periodical inspection ...

Page 6: ...e it is dangerous and not recommended Do not use the power cord if it is bundled or tangled Do not use the power cord if it is bundled or tangled If it is used in this manner it can overheat and fire may be caused Do not process bend wring or stretch the power cord forcibly Do not process bend wring or stretch the power cord forcibly Fire or electrical shock may result Substances that can not be u...

Page 7: ...electric shock may be caused Do not use a branching receptacle which may cause the heat generation 2 Choose a proper place for installation Do not install this unit in a place where Rough or dirty surface Flammable gas or corrosive gas is generated Ambient temperature exceeds 35 C Ambient temperature fluctuates violently There is direct sunlight There is excessive humidity and dust There is a cons...

Page 8: ...when the power switch is turned ON or OFF and fire explosion may result Never use explosive substances flammable substances and substances that include explosive or flammable ingredients in this unit Explosion or fire may occur 4 When using a wet material Drain off well the water of wet material before putting it in chamber in case to dry it Otherwise it may cause dew condensation or excessively h...

Page 9: ...expectible trouble or malfunction CAUTION 7 Do not make an overload 8 Do not set samples in close formation The withstand load of shelf is 15kg uniform load Set the samples apart each other The temperature in furnace cannot be controlled if too much samples are set there Make sure to use the shelf and set samples apart each other so as to make the free space of 30 or more to the furnace to acquire...

Page 10: ...s down or moves by the earthquake and the impact etc To prevent take measures that the unit cannot fall down and not install to busy place Touching the unit may cause a burn during and just after the operation To prevent take measures that putting up a notice of operating etc 11 Setting of the shelf and sample Two shelves are attached on the product Set them in place in furnace correctly The tempe...

Page 11: ...y melt and cause a fire or electrical shock If the power cord becomes damaged wiring exposed breakage etc immediately turn off the power at the rear of this unit and shut off the main supply power Then contact your nearest dealer for replacement of the power cord Leaving it may cause a fire or electrical shock Connect the power plug to the receptacle which is supplied appropriate power and voltage...

Page 12: ...9 Description and Function of Each Part Main Unit Door handle Front view Rear view Main switch Earth leakage breaker Door Exhausted opening Production plate Control panel Power cord ...

Page 13: ...ile setting quick auto stop timer or auto stop timer Lights while quick auto stop timer or auto stop timer is running AUTO START Lamp Blinks while setting auto start timer Lights while auto start timer is running FIXED TEMP Lamp Blinks while setting fixed temperature operation Lights while fixed temperature operation is running PROGRAM Lamp Blinks while setting program operation Lights while progr...

Page 14: ...ogram Operation Sv 1 Program Temperature Setting Used for setting the temperature for each step in the program Sv 1 to Sv 30 is shown t 1 Program Time Setting Used for setting the time for each step in the program t 1 to t 30 is shown PS 3 Step Number to be Repeated Used for choosing the step number to be repeated under the program operation with repeat function Refer to Page 26 Use program repeat...

Page 15: ...d temperature operation Pressing the TIMER key displays AStp The setting temperature SV can be set by pressing the ENTER key The operation time tim can be set by pressing it again Pressing the START STOP key starts the auto stop operation 18 4 Auto Start Operation This operation is used to specify the period up to automatic start after power on Pressing the TIMER key displays AStr The setting temp...

Page 16: ... again when the POWER switch is pressed again manual reset 2 Calibration offset function This calibration offset function is for calibrating the difference occurred between the required in furnace temperature and control temperature sensor temperature of the controller This unit can be calibrated toward either plus side or minus side of the whole temperature range 29 3 Overheating prevention tempe...

Page 17: ...14 Operation Method Operation Mode Function Setting Key and Characters The operation mode setting and function setting use the key operation and characters show in the following figure ...

Page 18: ...current temperature in furnace operation mode character and setting temperature of overheating prevention device are displayed on respective screens 2 Set the temperature for overheating prevention Press the SUBMENU key Press the several times to select the setting character of overheating prevention temperature OH Press the ENTER key The current setting temperature is displayed with blinking on t...

Page 19: ...e setting Also it displays the current setting temperature with blinking The FIXED TEMP lamp blinks too Set the temperature by pressing the 4 Start operation Press the orange START STOP key for about one second The unit starts operation and the blinking FIXED TEMP lamp lights on 5 Stop operation Press the orange START STOP key for about one second The unit stops operation and the FIXED TEMP lamp l...

Page 20: ...justment is needed 2 Start timer operation Press the START STOP key for one second after deciding the time Timer operation starts with the FIXED TEMP and AUTO STOP lamps lighting on The timer is activated at the point when the START STOP key is pressed 3 Stop terminate timer operation The operation stops automatically at setting time Buzzer continues to sound for about five minutes at operation st...

Page 21: ...s the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting temperature display screen displays the current setting time with blinking Set the time using the Timer function The maximum setting time is 999 hours and 50 minutes The time can be set in increments of a minute under 99 hours and 59 minutes It can be set in inc...

Page 22: ...th the FIXED TEMP and AUTO STOP lamps lighting on Press the START STOP key to terminate the timer operation mode The screen returns to the initial setting screen To correct or check setting Changing the setting temperature or time during operation is possible by pressing the TIMER key Use the to change the setting value Press the ENTER key respectively after changing the setting Press the to displ...

Page 23: ...UTO START lamp blinks too Set the temperature using the Press the ENTER key again The measurement temperature display screen displays the character tim which indicates the timer setting The setting temperature display screen displays the current setting time with blinking Set the time using the Timer function The maximum setting time is 999 hours and 50 minutes The time can be set in increments of...

Page 24: ...check setting Changing the setting temperature or time during operation is possible by pressing the TIMER key Use the to change the setting value Press the ENTER key respectively after changing the setting They are not changeable after the unit starts operation In this case stop the operation by pressing the START STOP key then set the value again Press the to display the setting temperature opera...

Page 25: ...emperature by 10 within 1 minute about 10 minutes is needed to increase or decrease temperature by 100 from current temperature Repeat function Repeat function is useful in case the operation uses the program repeating the same program steps Refer to page 26 for the function Check if the controller has sufficient free pattern for the number of steps to be created The steps however using the repeat...

Page 26: ...r used and setting temperature of overheating prevention device 2 Select program mode program pattern Press the PROGRAM key once The measurement temperature display screen displays the previous program mode Press the PROGRAM key again to display the next program mode PrG1 PrG2 PrG3 Select the mode and press the ENTER key When PrG1 is selected the measurement temperature display screen displays End...

Page 27: ...mperature rise fall capability of unit For example about 50 to 60 minutes is needed to increase the temperature from room temperature to 600 for DR200 When the current temperature is supposed to be 25 it takes about 1 minute to increase the temperature by 10 accordingly takes about 60 minute until it reaches to 600 Add an extra considering the temperature stability time Therefore the time is 70 in...

Page 28: ...be set in increments of a minute under 99 hours and 59 minutes It can be set in increment of ten minutes over 100 hours The can change the setting time quickly when it is pressed continuously Press them discontinuously when fine adjustment is needed To correct or check setting Press the FIXED TEMP key to correct the created program or to check the setting value The screen returns to the former one...

Page 29: ...emperature screen indicates PS 7 in the example because repeat function is used at the seventh step The step number 1 to 7 can be input in the setting temperature display screen Enter the number 1 in the example using the Press the SUBMENU key The measurement temperature screen displays the character Pc n which indicates the repeating times Enter the value of repeating times 2 in the example with ...

Page 30: ...n Method Program Operation Programming Preparation Form 1 Please use this form by making copies Register with PrG1 PrG2 PrG3 PAt1 PAt2 PAt3 No Project Name Date Programmer Program Pattern 700 600 500 400 300 STEP ...

Page 31: ... PAt1 PAt2 PAt3 No Project Name Date Programmer Input Value Temperature Time min Repeat Function Step 1 To Times Step 2 Step 3 Step 4 Step 5 Step 6 Step 7 Step 8 Step 9 Step 10 Step 11 Step 12 Step 13 Step 14 Step 15 Step 16 Step 17 Step 18 Step 19 Step 20 Step 21 Step 22 Step 23 Step 24 Step 25 Step 26 Step 27 Step 28 Step 29 Step 30 ...

Page 32: ...ature display screen falls by the setting temperature while the temperature on furnace rises When it is set to the minus side the temperature on the measurement temperature display screen rises by the setting temperature while the temperature on furnace falls The unit has two point correction function which performs offset between low temperature zone and high temperature zone Please consult our l...

Page 33: ...ill cause injury if fall During a thunder storm During a thunderstorm turn off the power key immediately then turn off the circuit breaker and the main power If this procedure is not followed fire or electrical shock may be caused When open close door Do not get close to the traveling range of door when opening or closing it It may hit your hands or head and result in an injury Keep door close dur...

Page 34: ...sample Recovering after power failure When power is supplied after a power failure the device automatically starts operation again with the same state as just before the power failure It is danger that the device starts unattached operation after a power failure We recommend for you to turn off the switch of this unit if a power failure occurs during operation Double stacking The unit becomes very...

Page 35: ...intenance after returning the temperature of this unit to the normal one Do not disassemble this unit CAUTION Use a well drained soft cloth to wipe dirt on this unit Do not use benzene thinner or cleanser for wiping Do not scrub this unit Deformation deterioration or color change may result in Monthly maintenance Check the earth leakage breaker function 1 Connect the power cord 2 Turn the breaker ...

Page 36: ...nit as possible and recycle the reusable parts considering to the environmental protection The feature components of this unit and materials used are listed below Component Name Material Exterior Parts Outer covering Steel plate melamine resin coating Furnace Stainless steel SUS304 Door packing Glass cloth Natural rubber Heat insulation material Ceramic fiber Glass wool Plates PET resin film Elect...

Page 37: ...hort circuit detection ALARM lamp lights on Er 02 appears Triac is in short circuit Make a call for service Heater disconnecting detection ALARM lamp lights on Er 03 appears Heater is disconnected Make a call for service Memory error ALARM lamp lights on Er 15 appears Failure in internal memory Make a call for service Internal communication error ALARM lamp lights on Er 17 appears Failure in inter...

Page 38: ...rature fluctuates during the operation Too much samples Wind from air conditioner directly blows The change of ambient temperature is remarkable Samples are too moist The power supply voltage is lower than the proper value In the case if the error other than listed above occurred turn off the power switch and primary power source immediately Contact the shop of your purchase or nearest Yamato Scie...

Page 39: ...ales office Check following items before contact Model Name of Product Production Number Purchase Date About Trouble in detail as possible Minimum Retention Period of Performance Parts for Repair The minimum retention period of performance parts for repair of this unit is 7 years after discontinuance of this unit The performance part for repair is the part that is required to maintain this unit Se...

Page 40: ... function Auto recovering after power failure Calibration offset Self diagnostic functions Failure of Sensor heater SSR memory internal communication temperature inputting circuit automatic overheating prevention device overheating prevention device measurement temperature Safety device Earth leakage breaker Overheating prevention device External dimensions W D H mm 520 443 612 Internal dimensions...

Page 41: ...Symbol Part name ELB Earth leakage breaker CONT Control board TM Terminal block PIO Display circuit board H1 2 Heater TH1 Sensor for control X1 Main relay TH2 Sensor for overheating prevention SSR Breakerless relay CT Current transformer ...

Page 42: ...4 display circuit board VS4 Yamato Scientific 1020000051 Tough card 300mm Yamato Scientific 1130000008 Main relay AHE1255 200V 240V Matsushita 2050000044 SSR TRS5225 Toho Denshi 2160000035 Power cord kit 2 0sq 3p Yamato Scientific LT00037454 Earth leakage breaker BJS1532N 15A Panasonic DN104 CT CTL 6 S H URD 2170010005 Door packing 1368 glass cloth Yamato Scientific 6080000002 ...

Page 43: ...ate Ammonium perchlorate and other perchlorate Potassium peroxide Sodium peroxide Barium peroxide and other inorganic peroxide Potassium nitrate Sodium nitrate Ammonium nitrate and other nitrate Sodium chlorite and other chlorites Calcium hypochlorite and other hypochlorites INFLAMMABLE LIQUID Ethyl ether Gasoline Acetaldehyde Propylene chloride Carbon disulfide and other flammable substances havi...

Page 44: ...at this document forbids Unexpected accidents or breakdown may result in Note The contents of this document may be changed in future without notice Any books with missing pages or disorderly binding may be replaced Instruction Manual for High Temperature Oven Model DR200 First Edition May 18 2017 Yamato Scientific America Inc 925 Walsh Avenue Santa Clara CA 95050 Phone 800 292 6286 408 235 7725 ht...