Operation Method
Program Operation
The example shown below explains the method of program registration using PrG3.
4. Register program
①
Select PrG3 referring to 3 mentioned above..
②
Input the number of steps, temperature and time for respective steps using
the program creation sheet.
③
Press the ENTER key. The PA
t1
is displayed with blinking.
("End" is displayed if PrG1 is selected. In this case, go to
⑥
)
④
Select the unused pattern from among Pat1, Pat2 and Pat3 using the "
▲
▼
".
⑤
Press the ENTER key. "End" is displayed and the step number "10" is
also displayed with blinking.
"End" is a character which indicates the total step number to be used. "9"
will be input here.
⑥
INPUT "9", which is the total step number to be used here, using the "
▲
▼
".
⑦
Press the ENTER key. The character "SV-1", which indicates the setting
temperature of the first step, is displayed. The current setting
temperature is also displayed with blinking.
⑧
Set the temperature of the first step using the "
▲▼
". "600" is input here
to set the temperature to 600
℃
.
⑨
Press the ENTER key. The character "t-1", which indicates the setting
time of the first step, is displayed. The current setting time is also
displayed with blinking.
Before setting the time, check the temperature rise/fall capability of unit.
For example, about 50 to 60 minutes is needed to increase the temperature
from room temperature to 600
℃
for DR200. When the current temperature
is supposed to be 25
℃
, it takes about 1 minute to increase the temperature
by 10
℃
, accordingly takes about 60 minute until it reaches to 600
℃
. Add an
extra considering the temperature stability time. (Therefore the time is "70" in
this example.)
The setting time of timer in respective steps is 999 hours and 50 minutes
maximum.
⑩
After the time is set, press the ENTER key.
⑪
The character "SV-2", which indicates the setting temperature of the
second step, is displayed. In the same way, input the temperature and
time for respective steps using the program creation sheet.
The different method is necessary where program repeat function is used.
In this case, press the SUBMENU key after setting the time (t-7 in the
example) in the step where the repeat operation is to be used (Step 7 in
the example). This enters to the repeat function setting mode.
Follow the "Use program repeat function" in page 26 for the input method of
program repeating function.
⑫
The screen returns to the initial setting screen after the setting of
temperature and time in the final step is completed.
Verification run:
Make sure to check the setting temperature and time by operating the unit
without load before performing actual run with samples.