31
Handling Precautions
Setting of sample
Since the withstand load of the attached shelf plate is about 15kg per one plate, do not set heavier
sample than 15kg. When setting several sample, set them as dispersed as possible. Too much
sample setting could cause the improper control of the temperature. For keeping the proper
temperature, keep more than 30% space against whole size of the shelf plate, and set the
sample.
Recovering after power failure
When power is supplied after a power failure, the device automatically starts operation again with
the same state as just before the power failure. It is danger that the device starts unattached
operation after a power failure. We recommend for you to turn off the switch of this unit if a
power failure occurs during operation.
Double stacking
The unit becomes very hot. Do not make double stacking.
After installing
It may cause injure to a person if this unit falls down or moves by the earthquake and the impact.
etc.. To prevent, take measures that the unit cannot fall down.