4. Operating procedure
Calibration offset
3 Set Calibration Offset value.
↓
①
Press
key to display [
CAL
:
OS
] on Bottom
Screen then press
key.
②
Input offset value by
and
keys
and then press
key. You can enter an
offset amount up to ±15.0
℃
Example
Read temperature
:
200
℃
and actual measured
temperature
:
198
℃
⇒
Offset input value: -2.0
℃
※
Although you can input values up to the first
decimal place, the temperature indications and
measured temperatures will be rounded before
indication.
③
Pressing the
key longer will return to the
time display screen.