22
4. Operating procedure
Fixed temperature operation
FIXED TEMP(Fixed Temperature) mode is to keep running at target temperature. It will keep running
until operation stops.
SV
: Set Value (Target Temperature), t
:
Time
Set Fixed Temperature mode.
1 Turn on the controller.
Turn on Earth Leakage Breaker(ELB) on
(|)
the right
side wall of this Equipment. (
Idle State
)
Press and hold key to turn on the controller power.
(
Standby State
)
Indicate read temperature in Chamber on Top Screen
and indicate target temperature on Bottom Screen.
The fan motor will start and FAN lamp iluminats.
The fan motor operates when the door is closed and
stops while the door is open.
2 Select Fixed Temperature Operation.
Press
key to turn FIXED TEMP (Fixed
Temperature mode) lamp on.
※
Fixed Temperature mode would be selected at first
time operation. After that, most latest operated
mode is selected.
3 Set target temperature.
①
Press
key. Then flash changeable digits on
Bottom Screen.
②
Shift to flashing digit with
key and then change
to desired digit with
keys.
Operating Temp. Range:0
~
270
℃
③
Press
key when target temperature setting has
completed.
Press
key once or twice to cancel its setting.
SV
t
▲
Stop operation (manually)
▲
Start operation (manually)