26
4. Operating procedure
Auto stop operation
This operation mode is used to automatically stop operation by setting the timer.
The operation mode where operation is automatically stopped by setting an operation duration.(when
you set an operation duration)
※
When you set a time, the
wait function will be activated,
the mode will remain “waiting”
without counting down the time
until temperature indication will
be within the wait deviation
range between -3
℃
and +6
℃
to
the
set
temperature.
Counting down starts when the
temperature in the chamber
reaches the temperature -3
℃
(indication)
to
the
set
temperature.
Even if the temperature in the
chamber (indication) the mode
will be “waiting” if the lower limit
of the wait width deviation is
exceeded and time counting
down will not occur until the
temperature in the chamber
(indication) returns.
Operation mode where operation stops automatically at the set time (when an operation time is set)
※
The wait function will not
work if you select a time
setting. Operation will stop
when the set time comes. The
time you can set is up to 24
hours from the present time.
When a power failure occurred
before the set time and
continued after that and then
the
unit
recovered
automatically, operation will
continue to the next set time so
remember to stop operation
manually.
Set Automatic Stop mode
1
Turn on the controller
Turn on
(|)
Earth Leakage Breaker (ELB) on the right
side wall of this Equipment. (
Idle State
)
Press and hold key to turn on the controller power.
Indicate circulating liquid temperature in Chamber on
Top Screen and indicate target temperature on Bottom
Screen.
The fan motor will start.
The fan motor operates when the door is closed and
stops while the door is open.
Temperature
Set
temperature
Lower limit of the wait
width deviation
Door open
Timer start
①
Timer start
②
Wait
①
Wait
②
Actual elapsed time=set time (timer start
①
+
②
)+(Wait
①
+
②
)
Operation start (manual)
Operation stop (automatic)
Time
Temperature
Set
temperature
Door open
Operation start (manual)
Operation stop (automatic)
Time
Set time