25
4. Operating procedure
Auto stop operation
2 Selecting Automatic stop Operation
Press
key to turn FIXED TEMP (Fixed
Temperature mode) and AUTO STOP (Automatic Stop
mode) lamp on.
※
Fixed Temperature mode would be selected at first
time operation. After that, the latest operated mode
is selected.
3 Set target temperature and operation
running time / clock time to stop.
①
Press
key.
Select stop method from TIME/CLOCK with
keys and then press
key.
②
Set TIME
(
capable setting range: 0
~
99hr
:
59min
)
or CLOCK (according to 24-hour time) on Top
Screen and then press
key.
③
Set target temperature on Bottom Screen and then
press
key.
Example 1.
Setting running time:
Operation is stopped automatically in 35 hours and 30
minutes once temperature reached to 250
℃
of target
temperature.
Example 2.
Setting clock time to stop:
Start operation and reach to 250
℃
in Chamber of
target temperature, and operation is stopped
automatically at 15:00.