2. Before operating the Equipment
Precautions when installing the Equipment
9. Connect Power Cord/Power Cable to receptacle or switch board of facilities.
Connect Power Cord/Power Cable to suitable receptacle/switch board of facilities according to
electrical requirements as follows.
Electrical
requirements:
DNF301
AC115V single phase 50/60Hz 8A or more
(
ELB capacity
;
15A
)
DNF401
AC100V single phase 50/60Hz 11.5A or more
(
ELB capacity
;
15A
)
DNF411
AC200V single phase 50/60Hz 7.5A or more
(
ELB capacity
;
10A
)
DNF601
AC100V single phase 50/60Hz 15.5A or more
(
ELB capacity
;
20A
)
DNF611
AC200V single phase 50/60Hz 10A or more
(
ELB capacity
;
15A
)
The operational voltage range is ±10%, the voltage range where the specified performance is
guaranteed is rating±5%, the frequency is rating±1%.
※
Check line voltage of its receptacle/switch board of facilities and/or whether utilize the same
line with other equipments or not, if this Equipment does not start up/operate even to turn Earth
Leakage Breaker(ELB) On(
|
). Take correct action for the solution, such as changing its power
source away from other equipment.
Multiple plugs with a branch outlet or extension with the cord reel may cause voltage drop,
resulting in deterioration of the heating capacity or of the temperature control performance.
10. Must connect grounding wire properly.
(
DNF301 DNF401specifications
)
・
Must connect grounding wire properly to grounding line or terminal in order to avoid
electrical shock due to electrical leakage.
・
Never connect grounding wire to gas line pipe or water line pipe due to fire or electrical
shock.
・
Never connect grounding wire to telephone grounding line or to lightening conductor due to
fire or electrical shock.
・
Never connect multiple plug to single receptacle due to generating heat dangerously.
Connect to grounded receptacle.
Receptacle with ground connection
When there is no ground terminal.
● Require to ground by Electrical Equipment
Technical Standards Section 19-calss D in
Japan. Please contact with local dealer, local
electrician, or Yamato Customer Service
Center.
Use grounded adaptor for ungrounded receptacle
Ground adaptor
● Insert grounded power plug into ground adaptor.
Connect grounding wire(green) of ground
adaptor to ground terminal on switch board of
facilities.
Ground-pole
Grounded 3-pole power plug
Ungrounded 2-pole
receptacle for 115V AC
Grounded
3-pole
power plug
Ground adopter
Grounding
wire