60
5. Handling precautions
Warnings and Cautions
Material
Classification
Silicon Rubber
Fluoro-rubber
Fatty Acid, Phenol
Acetic Anhydride, Oleic Acid,
Phenol Palmitate
Formic Acid
、
Acetic Anhydride,
Hydroquinone
Nitrogen Chemical
Compounds
Nitromethane, Nitroethane,
Nitropropane
Nitromethane, Nitroethane,
Nitropropane, Ethylenediamine,
Dimethylaniline, Ethanol amine,
Hydrazine, Triethanol Amine,
Dimethyl Formamide, Pyridine,
Piperidine
Sulfur and
phosphorus
compounds
Hydrosulfuric
Hydrosulfuric,
Tributyl Phosphate
Other Chemical
Compounds
Nickel Acetate, Lead Acetate,
Zinc Acetate, Tetraethyl Lead,
Vegetable Oil, Silicon Oil
Calcium Acetate, Nickel Acetate,
Lead Acetate, Zinc Acetate
Inorganic Solvent
Hydrochloric Acid, Nitric Acid,
Sulfuric Acid, Hydrobromic Acid,
Phosphoric Acid, Hypochlorous
Acid,
Chromic Acid, Perchloric Acid,
Sodium Hydrate
Sodium Hydrate,
Aqueous Ammonia
22. Never fail to perform periodic inspection.
Check regularly Earth Leakage Breaker (ELB) and Independent Overheat Prevention Device
(IOPD) which they are key part/Device for the safety of this Equipment.
Refer to Chapter 6. Maintenance Method on page 61.
23. Pay attention to the following when operating the product with the exhaust
damper fully opened.
With the exhaust damper fully opened, the set temperature cnnot be reached.
Note also that the set temperature may not be reached in cases other than the dampler fully
opened. The customer is requested to check this in line with the variable speed before
operation.
24. Take care for possible degradation of performance when using the cable port.
When a measurement sensor or a probe is inserted into the cable port close the cable port
cover as much as possible and completely seal to any gaps with heat-resistant packing or
sealant. If seal is insufficient, the temperature characteristic, cleanliness or other performance
will degrade.
Use an optional silicon plu
g (with one hole) as necessary. See”P.70 List of optional settings”.
25. Smoke (odor) may generate when you operate the unit for the first time.
When you operate the unit for the first time, the bonding material of the heat insulation material
may burn and generate odor, which, however, does not indicate a malfunction of the unit. Odor
will not generate as you continue to use the unit for some time.