4. Operating procedure
Programming Method
Sample program setting In this example, 8 steps are registered in the program pattern 2, steps
from 4 to 7 will be repeated 5 times and the whole session will end at
the step 8.
Note: Steps 4 to 7 will be repeated 6 times.
Pattern
No
Step
Set
temp.
Set
time
Repeat
dstn.
Repeat No.
Wait
Wind
speed in
operation
End
Wind
speed
at end
P
*
*
:01
P02:
**
TEMP
TIME
REP(
STEP
)
REP(
COUNT
)
WAIT
FAN
ENDST
FAN
02
01
100
00:00
0
0
ON
10
OFF
-
02
100
01:15
0
0
OFF
10
OFF
-
03
150
00:00
0
0
ON
10
OFF
-
04
150
02:00
0
0
OFF
10
OFF
-
05
200
01:00
0
0
ON
10
OFF
-
06
200
01:30
0
0
OFF
10
OFF
-
07
100
00:00
4
5
ON
0
OFF
-
08
100
00:30
0
0
OFF
0
ON
0
※
When time settings on heating or cooling steps are beyond the heating or cooling capacity (0
minutes in steps 1, 3 & 7 above) of the unit, it will operate at full power for a short time in wait (
ON
)
mode until temperature setting has been reached. With wait set to
OFF
, unit will proceed to the
next step regardless of whether temperature setting has been reached. Use caution when setting
short heating/cooling times.
※
When the time setting on heating or cooling steps is set longer than unit normally takes build heat
or cool, unit will adjust itself to do so within the set timeframe.
※
When a fixed temperature step is set and wait is [
ON
], the wait mode will continue from the time
when the temperature in the bath drops below the lower limit of the wait width deviation temperature
due to, for example, opening of the door until the temperature in the bath will recover above that
lower limit. At [
OFF
] the process will proceed to the next step after the set time irrespective of
changes of the temperature in the bath.
※
When you use the repeat function, program the operation so that the set temperature before
shifting to the repeat mode will be the same as the set temperature of the destination of repetition.
※
Checking the heating capacity and the cooling capacity before setting is encouraged since these
will differ depending on the environmental temperature and the operating conditions.
※
FAN: F
100
150
200
STEP1
1時間15分
STEP2
0分
STEP3
0分
STEP4
2時間
リピート回数:5回
STEP5
1時間
STEP8
1時間30分
STEP7
0分
30分
温度
時間
(℃)
STEP1 3 7:温度上昇
/
下降を最短時間で設定
STEP5:温度上昇を定率上昇時間で設定
STEP6
0 min.
0 min.
1 h. 15 min.
1 h.
1 h. 15 min.
Setting heating/cooling in minimum time
Setting heating in a fixed rate heating time
Time
Repeat five times
Time
Temp.
0 min.
30 min.
2 h.