10. SPECIFICATIONS
Product
Clean oven
Model
DE411
DE611
DT411
DT611
System
Forced air circulation and ventilation
External temperature range
5
℃~
35
℃
Power supply
Triple phase
220V AC 7A
Triple phase
220V AC10A
Triple phase
220V AC 10A
Triple phase
220V AC 14A
Common to 50/60Hz, operating voltage range
:
±10
%
P
erfor
m
an
c
e
*1
Temperature Range
Room temp. + 30
℃~
260
℃
Room temp. + 30
℃~
360
℃
Temperature control
accuracy
±0.3
℃ (
at 260
℃)
JTM K05
±0.3
℃ (
at 360
℃)
JTM K05
Temperature
fluctuation
*2
±0.5
℃ (
at 260
℃)
JIS C60068
±0.5
℃ (
at 360
℃)
JIS C60068
Temperature
distribution accuracy
±2.5
℃ (
at 260
℃)
JTM K05
±4.0
℃ (
at 360
℃)
JTM K05
Temperature slope
10
℃ (
at 260
℃)
JIS C60068
20
℃ (
at 360
℃)
JIS C60068
Time to max temp.
Approx. 70 min.
Approx. 80 min.
Clean level
Class100 when temperature stable *3
Conf
igu
ra
ti
o
n
Exterior
Chrome-free electro-galvanized steel
Chemical-proof powder coating
Chamber
Stainless steel
Insulation Material
Glass wool
Door
Single side (left) swing
Heater
Stainless steel tube heater with fins
Heater capacity
2.5kW
3.6kW
3.6kW
5.2kW
Fan (motor)
Stainless steel sirocco fan (condenser motor 400W)
Differential gauge
Analog type colored scale
(
0
~
300Pa
)
Cable port
I.D. :φ33mm
(
right side
)
Exhaust port
Manual exhaust damper
O.D.φ61mm
(
top panel
)
Casters
Universal swivel caster (no stoppers)
Adjustable leveling
feet
Two (front undercarriage)
HEPA filter
Heat-resitant HEPA filter
(
dust-collection efficiency
:
99.97% or better, up to
0.3μm particle filtering
)
Intake filter
Stainless steel felt
Cont
ro
l Co
nfi
gu
rat
io
n
Type
V-shape controller
Temp. control
system
PID Z control
Temp. setting
system
Digital setting with ▲/▼ keys.
Temp. display
system
Upper display (Chamber): Green 4-digit LED Digital Display (Incrementation:
1
℃
)
Lower display: Orange 5-digit LED Digital Display (Incrementation 1
℃)
Other displays
LED: temperature indicator for heating/stabilization/cooling
Timer
Configurable from 1 minute, 99 hours, 59 minutes,
timer operation:
24 hour time system, clock operation