19
4. OPERATION PROCEDURE
Auto Stop Operation
AUTO STOP (Automatic Stop) utilizes timer or clock to automatically stop an operation. Operation
must be started manually. See below.
Auto stop mode set to timer:
When auto stop mode is set to
the timer, unit enters a wait
period, and remains
“waiting”
without counting down time
until chamber temperature is
within a set deviation range
(-3°C ~ +6°C) of the
temperature setting.
Countdown begins once
chamber temperature is within
the deviation range.
If temperature drops below or
exceeds the deviation range
afer stabilzing, such as when
door is opened, allowing heat
to escape; countdown stops
and unit once again enters
wait period until heat builds to
within deviation range once
again.
Auto stop mode set to clock:
Wait function does not operate
when auto stop mode is set to
the clock. Operation stops as
soon as clock and set time
agree, regardless of when
temperature setting is
reached.
If a power failure occurs while
auto stop mode is set to clock,
unit will recover operation
automatically when power is
restored and run until stop
time.
Set auto stop mode
1
Turn on power
Turn on (
|
) main power switch (ELB) (idle).
Press and hold to turn on control power (standby).
Current temperature reading is shown in upper display,
Temperature setting is shown in lower display.
Fan begins running (runs when door is closed and
stops whenever door is opened).