Turb PLUS 2000
Installation
16
100193 Rev 0 ba77254e01 11/2020
Figure 4: Instrument Parts Identification
Water Inlet Tubing
Connection
Use ¼” OD tubing with
supplied
push-to-connect fitting.
Polyethylene or
polypropylene are
recommended tubing types.
Backpressure Valve
This valve allows for an
adjustment to reduce bubble
formation. Turn clockwise to
increase backpressure.
Shutoff Clamp
This clamp allows for shutting
off the intake flow during
cuvette maintenance. Press
down to restrict water flow.
Drain Tubing Connection
Use ¼” OD tubing with supplied
push-to-connect fitting. Do not
reintroduce drain sample back
into the process stream.
Pressure Regulator
The integral pressure regulator is
rated for a maximum pressure
of
7 bar (101.5 PSI). The maximum
allowable fluid
temperature is
50°C (122°F).
Flow Through Cuvette
The flow through cuvette is
rated for
a flow of 100ml/min -
1liter/min. (0.026-0.26Gal/min).