Installation
Turb PLUS 2000
100193 Rev 0 ba77254e01 11/2020
17
5.3.3
Connecting the drain
CAUTION
Turb PLUS 2000 drain tubing MUST be routed to a suitable sanitary
sewer drain. Wetted materials are not FDA approved.
DO NOT return the drain sample or any water run through Turb PLUS
2000 into the process stream or into any potable water supply, as this
results in contamination of drinking water.
1. Select a length of ¼ inch OD flexible tubing (Polyethylene or polypropylene) long enough to directly
connect the Turb PLUS 2000 drain connection point to a suitable sanitary sewer drain. Avoid
unnecessary bends or sags in the tubing. Make sure there are no kinks, twists, breaks, or damage of
any kind in the tubing.
2. Using a sharp utility knife, cut the end of the tubing square. Remove any burrs or sharp edges before
inserting into fitting.
3. Place a piece of tape ½ inch from the end of the tube to indicate how far the tube should be inserted.
4. Push the tube into the Quick-Connect fitting on the right facing side of the Turb PLUS 2000, as shown
in Figure 4.
5. Check to confirm that tube is pushed fully against the tube stop. Gently but firmly pull on the tube to
check that it is secure.
6. Connect the other end of the tube to a suitable sanitary drain or trench following proper plumbing
techniques applicable to your facility and configuration.
5.3.4
Connecting the water intake
CAUTION
Use Turb PLUS 2000 ONLY with water. Connect water intake ONLY to
water free of corrosive chemicals or other hazardous substances.
Xylem Analytics Germany accepts no responsibility for damage caused by the
introduction of vapors, fluids or other materials into the instrument process
stream which is not compatible with the instrument's wetted materials. A list of
the wetted materials can be found in the specifications on page 2 of this
manual.
1. Select a length of ¼ inch OD flexible tubing (Polyethylene or polypropylene) long enough to directly
connect the Turb PLUS 2000 water inlet connection point to water sampling point. Avoid unnecessary
bends or sags in the tubing. Make sure there are no kinks, twists, breaks, or damage of any kind in the
tubing.
2. Using a sharp utility knife, cut the end of the tubing square. Remove any burrs or sharp edges before
inserting into fitting.
3. Place a piece of tape ½ inch from the end of the tube to indicate how far the tube should be inserted.
4. Push the tube into the Quick-Connect fitting on the left facing side of the Turb PLUS 2000, as shown in
Figure 4.
5. Check to confirm that tube is pushed fully against the tube stop. Gently but firmly pull on the tube to
check that it is secure.