Installation and Commissioning
Operating environment
2.1
Observe the following points to ensure that the device functions properly and has a long service life.
The ambient temperature must be10 to 40 °C.
The relative humidity must be below 80 % (without condensation).
The titration head of the sample changer must be able to move freely.
Do not use or store the device in dusty, damp or wet places.
Make sure that no liquids can get into the device and wipe off any liquids on the device immediately.
Protect the device from vibrations, direct sunlight, corrosive gases and strong magnetic and/or electric
fields.
Use only original spare parts and accessories.
Unpacking and setting up
2.2
The devices have been individually assembled for you, therefore there may be deviations with regard to the
described scope of delivery and the accessories. Please refer to the enclosed packing list for the exact scope of
delivery. If you have any questions, please contact us directly (for the service address,
see backside of this
operating manual
).
The devices and all accessories as well as the peripheral devices have been carefully checked at the factory for
function and dimensional accuracy. Please ensure that even the small additional parts are removed from the
packaging without leaving any residue.
The units can be set up on any flat surface. Do not push any objects under the devices.
For the operation of the TO 7280 / TW 7650 an additional coulometer module TitroLine
®
7500 KF
trace
is
necessary. The TitroLine
®
7500 KF
trace
is already included in the scope of delivery of the TM 235 TO M6.
For the scope of delivery of the TitroLine
®
7500 KF
trace
see the according operating manual.