52
1.3.3
Titration stand TM 235 TO
Translation of the legally binding German version
(Release: 22. April 2021)
In combination with the titrator TitroLine
®
7500 KF
trace
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: for laboratory equipment
RoHS Council Directive 2011/65/EU
Country of origin
: Germany, Made in Germany
Pump
:
Free flow rate - air externally adjustable between 50
… 500 ml / min
Maximum pressure: 1.5 bar
Valve
:
Automatically switchable valve for switching between air and gas flow
Control
:
Via USB connection of the titrator TitroLine
®
7500 KF
trace
Stirring Speed
:
50 ... 1000 U/min, adjustable via a control dial
Gas flow meter
:
Adjustable between
50 … 500 ml / min
Connections
Power supply:
Low voltage input 12 V /
on the backside of titration stand
Plug connection: plug for low voltage connection
– phone jack-,
Positive pole at pin contact, inside contact
= 2.1 mm, USA/Japan,
Power supply via titrator TitroLine
®
7500 KF
trace
USB
:
For connection to the TitroLine
®
7500 KF
trace
(see control unit)
Housing
:
Material:
Polypropylen
Dimensions:
8 x 13 x 25 cm (W x H x D), Height without stand, bottle and titration vessel
Weight:
ca. 1.9 kg
Ambient conditions
:
Not suitable for use in explosive environments!
Climate:
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
Altitude:
No restrictions
Pollution degree: Pollution degree IP 20, indoor use only.