50
Technical Specifications
1.3
1.3.1
Headspace Oven TO 7280
Translation of the legally binding German version
(Release: 22. April 2021)
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: for laboratory equipment
RoHS Council Directive 2011/65/EU
Country of origin
: Germany, Made in Germany
Measuring range
: (
μg water absolute) in combination with a TitroLine
®
7500 KF
trace
Coulometer:
Measuring range
Resolution [µg]
Reproducibility
1 µg
… 200 mg
0.1
+/- 0.3
g at 10
– 1000µg
0.3 % at > 1000 µg
Temperature range
:
40
°C … 280 °C (isotherm)
Power supply
:
by external multi-range power supply from 115
– 230 V, 50/60 Hz
Input voltage: 24 Volt DC, 11000 mA
Power consumption 250 W
Corresponds to protection class III:
Only use the power supply included in the scope of delivery or a power supply
approved by the manufacturer
!
Display
:
3 colored LED display (blue/red/green).
Operating states, error states and operating faults are signaled by color signals
and flashing codes
RS-232-C Interface
:
Daisy chain function available
RS-232-1
for connection to a titrator, Daisy Chain input
Data bits: fixed
8
Bit
Stop bit:
fixed
1
Bit
Parity:
fixed
no
Baud rate: fixed
4800
Baud
Address: fixed
3
RS-232-2
for connection of balances, Daisy Chain output
Data bit:
configurable, 7 or
8
Bit (Default Wert 8 Bit)
Stop bit:
fixed
1
Bit
Parity:
configurable: even / odd /
no
(Default no)
Baud rate: configurable: 1200, 2400,
4800
, 9600, 19200, 38400 (Default 4800 Baud)
Housing
:
Material:
Powder coated aluminum, steel, stainless steel
Dimensions:
30 x 45 x 24 cm (W x H x D)
Weight:
approx. 7 kg for basic unit without titrator and titration stand TM 235 TO