Calibration 

Allow at least 5 minutes warm-up before commencing calibration. 

Access to Calibration Adjustments 

All adjustments are on the front panel control board; all the trimmers are adjusted from the 
reverse side of the board through holes in the board.  To gain access to the board it is necessary 
to remove the top cover. 

                                                               WARNING! 

When the instrument is connected to its supply the removal of covers is likely to expose live parts. 
The instrument should be disconnected from all voltage sources before it is opened for any 
adjustment, replacement, maintenance or repair. Capacitors inside the power supply may still be 
charged even if the power supply has been disconnected from all voltage sources but will be 
safely discharged about 5 minutes after switching off power. 
Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided 
as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of 
the hazard involved. 
Remove the 6 side screws and the front handle screw to release the top cover. 
Component adjustment references are given as Channel A/ Channel B, e.g. VR5/VR105; 
Channel A is the left-hand output. 

                           Control pcb calibration points (viewed from rear) 

Equipment Required 

A 5½ digit multimeter with better than 0.05% accuracy on dc volts and better than 0.1% accuracy 
on dc current (to 4A); alternatively use a precision shunt for current measurement. 

Rheostat or other high power load arrangement to provide up to 4A load at 75V. 

Mode A (75V/2A x 2) 

Adjust Channel A and Channel B independently as follows. 

Voltage Calibration 

Connect the DMM (set to Volts) across the output. Set voltage and current controls to minimum. 
Switch output ON (Check LED is on) and check for a reading of 00.0V ± 0.1V on the Volts display 
and DMM; check the Amps display reads 0.00 ± 0.01A. 

Set voltage and current controls to maximum. Adjust VR5/VR105 (maximum output volts) for a 
reading of 75.35V to 75.45V on the DMM. Adjust VR6/VR106 (measured output volts) until the 
Volts display matches the reading on the external DMM. 
Switch output OFF. Adjust VR10/110 (preset volts) until the Volts display shows 75.4. 

 10 

Summary of Contents for XPH 75-2D

Page 1: ...XPH Series Dual 75V 2A DC Power Supply XPH 75 2D Operating Manual...

Page 2: ...e the property of their respective owners and are used herein for identification purposes only Notice of Copyright Xantrex International All rights reserved Disclaimer UNLESS SPECIFICALLY AGREED TO IN...

Page 3: ...ion 11 Connexions 11 Fonctionnement 11 Maintenance 13 Bedienungsanleitung auf Deutsch Sicherheit 14 Installation 15 Anschl sse 15 Betrieb 15 Wartung 17 Istruzioni in Italiano Sicurezza 18 Installazion...

Page 4: ...d good transient response High power efficiency ensures that the unit remains cool without any fan noise The XPH 75 2D has been designed to meet the stringent requirements of relevant IEC standards fo...

Page 5: ...ding rate 4 Hz Meter Resolutions 100mV 10mA Meter Accuracies Voltage 0 3 of reading 1 digit Current 0 6 of reading 1 digit Operating Modes Output 1 meters are blanked in Modes B and C GENERAL AC Input...

Page 6: ...ted by the use of an extension cord without a protective conductor When the instrument is connected to its supply terminals may be live and opening the covers or removal of parts except those to which...

Page 7: ...ormance A According to EN61326 the definitions of performance criteria are Performance criterion A During test normal performance within the specification limits Performance criterion B During test te...

Page 8: ...BE EARTHED Any interruption of the mains earth conductor inside or outside the instrument will make the instrument dangerous Intentional interruption is prohibited Connections All connections are mad...

Page 9: ...he terminals of Output 1 are isolated from those of Output 2 Note that with the output switch of Output 1 in the ON position the Output lamp still lights but the terminals remain open circuit Note als...

Page 10: ...ide a current equal to the sum of the current limit settings Ventilation The power supply is very efficient but nevertheless can generate significant heat at full power The supply relies on convection...

Page 11: ...hazard involved Remove the 6 side screws and the front handle screw to release the top cover Component adjustment references are given as Channel A Channel B e g VR5 VR105 Channel A is the left hand o...

Page 12: ...current control to maximum and adjust load for 2A output current Note that the DMM should be connected at the back of the output terminals closest to the internal sense connection and the load plugge...

Page 13: ...strument est reli son alimentation il est possible que les bornes soient sous tension et par suite l ouverture des couvercles ou la d pose de pi ces l exception de celles auxquelles on peut acc der ma...

Page 14: ...connexions sont effectu es au panneau avant La charge doit tre reli e aux bornes positive rouge et n gative noire marqu es OUTPUT Sortie Relier une sortie ou aux deux selon les besoins pour le mode s...

Page 15: ...commande de limite de courant doublent de valeur lorsqu on passe du mode A au mode B c d que lorsque la commande est r gl e une limite de 1 5 A en mode A la limite passe 3 A en mode B Pour viter d end...

Page 16: ...n refroidissement par convection uniquement et il est donc important que la ventilation ne soit jamais r duite afin d assurer une bonne performance et s curit Maintenance Le Constructeur ou ses agents...

Page 17: ...abels ohne Schutzleiter nicht aufgehoben werden Ist das Ger t an die elektrische Versorgung angeschlossen so k nnen die Klemmen unter Spannung stehen was bedeutet da beim Entfernen von Verkleidungs od...

Page 18: ...schlu an an einen oder beide Ausg nge vor dem gew hlten Modus entsprechend Siehe hierzu Betriebsabschnitt Die mit dem Symbol gekennzeichnete Klemme ist mit dem Chassis und der Schutzerde verbunden Bet...

Page 19: ...s zu vermeiden sind vor dem Umschalten die Ausg nge auszuschalten und die Strombegrenzung ist auf den gew nschten Wert einzustellen bevor der Ausgang wieder eingeschaltet wird Modus C Befindet sich de...

Page 20: ...ie Ventilation niemals eingeschr nkt wird wenn Leistung und Sicherheit aufrecht erhalten werden sollen Wartung Die Hersteller bzw deren Vertretungen im Ausland bieten die Reparatur von Ger ten an bei...

Page 21: ...di una prolunga priva conduttore di protezione Quando lo strumento collegato all alimentazione alcuni morsetti sono sotto tensione e l apertura dei coperchi o la rimozione di parti eccetto quei compo...

Page 22: ...onda della modalit selezionata Vedere la sezione Funzionamento Il morsetto segnato collegato allo chassis e al terminale di sicurezza di terra Funzionamento In modalit A indipendente il funzionamento...

Page 23: ...ova le uscite devono essere spente prima di cambiare la modalit e il limite di corrente deve essere impostato sul valore voluto prima di riaccendere l uscita Modalit C Con il commutatore in questa pos...

Page 24: ...ne non sia mai ristretta se si vogliono mantenere i livelli ottimali di rendimento e sicurezza Se il dispositivo montato in uno spazio ristretto ad es su una rastrelliera da 19 pollici si deve provved...

Page 25: ...arse por el uso de una extensi n de cable sin conductor protector Cuando el instrumento est conectado a su suministro es posible que queden sin protecci n elementos bajo tensi n y la abertura de tapas...

Page 26: ...el panel delantero La carga se ha de conectar a los terminales positivo rojo y negativo negro marcados OUTPUT Salida Conecte a una o ambas salidas seg n corresponda al modo seleccionado ver el apartad...

Page 27: ...uando se cambia de Modo A a Modo B es decir ajustando el control a un l mite de 1 5A en el Modo A el l mite se convierte en 3A en el Modo B Para evitar posibles da os al circuito que se prueba las sal...

Page 28: ...ntilaci n El suministro de fuerza es muy eficaz pero puede generar mucho calor a fuerza plena El suministro depende de enfriamiento por convecci n solamente y por lo tanto es importante que la ventila...

Page 29: ...and return shipment to the customer via a Xantrex selected non expedited surface freight within the contiguous United States and Canada Alaska and Hawaii are excluded Contact Xantrex Customer Service...

Page 30: ...HANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE TO THE EXTENT REQUIRED UNDER APPLICABLE LAW TO APPLY TO THE PRODUCT SHALL BE LIMITED IN DURATION TO THE PERIOD STIPULATED UNDER THIS LIMITED WARRANTY IN...

Page 31: ...reached during work hours A brief description of the problem 3 Ship the unit prepaid to the address provided by your Xantrex customer service representative If you are returning a product from outside...

Page 32: ...chnology Inc 8999 Nelson Way Burnaby British Columbia Canada V5A 4B5 604 422 8595 Tel 604 421 3056 Fax 800 667 8422 Toll Free North America prg info xantrex com www xantrex com 975 0102 01 02 48511 07...

Reviews: