- 8 -
Temperature Calibration Offset
1. Turn on the power switch.
2. Ensure the desoldering gun is in the holder.
3. Press the * button three times. The LCD will read “R00”. The “00”
indicates that the current calibration offset temperature is 0°.
4. Press the increase/decrease button to set the desired temperature
offset.
5. The unit will automatically save the setting and exit the function
menu after six seconds.
6. The Temperature calibration offset has now been set.
Temperature Conversion
1. Turn on the power switch.
2. Ensure the desoldering gun is in the holder.
3. Press the * button four times. The LCD will read “C”. This indiates
that the current temperature measurment is in °C.
4. Press the increase/decrease button to set the desired temperature
output mode.
5. The unit will automatically save the setting and exit the function
menu after six seconds.
6. The will now display the temperature in the desired unit of
measure.