Error Messages
Errors encountered during a measurement are displayed on the
instrument screen. All errors are accompanied by a long beep
and flashing yellow light. The error message is cleared from the
instrument screen by pressing the Enter key
#
.
Displayed Errors: Reason
Measurement was
aborted by user
Displays with an incomplete measurement or
calibration. Instrument was released too
soon.
Calibration has
timed out
The calibration interval time set in the
configuration has been reached. Calibration
is now required.
Calibration
required
Displays anytime the instrument requires a
calibration.
Calibration has
failed
Calibration failed. Make sure the instrument
is properly positioned on the reference.
The battery is
getting low
This warning appears when the battery falls
below approximately 25% of full charge.
Measurements are still possible, but the
battery should be charged soon.
Batteries are
dead
Displays when not enough battery power
remains to take measurement. The current
measurement is aborted.
Batteries are
missing
The battery pack is not installed. Instrument
will not allow any measurements.
Incorrect
Charger Voltage
Wrong charger is connected or charger is
bad.
The battery is
over–charged
The battery pack is too hot. Remove battery
pack and let it cool off.
Lamp is getting
weak, replace
soon
Reading lamp is at 50% strength or less from
its original intensity. Measurements are still
possible but the lamp should be replaced
soon.