MAINTENANCE
4-2
Calibrating the Instrument
The ColorEye XTH should be calibrated with the White Ceramic Calibration Standard (tile)
provided with the instrument in the pocket of the carrying strap. It is best to calibrate the
instrument at the location it is to be used. This will accommodate ambient temperature
changes.
Before You Calibrate
…
Please note the following before you calibrate the spectrophotometer:
•
If the instrument is exposed to rapid changes in ambient temperature, calibrate to
accommodate for these changes.
•
Make certain that the serial number on the White Ceramic Calibration Standard is
the same as the serial number on the instrument. Do not substitute another
calibration standard for the one originally supplied.
•
If the standard should be broken or become damaged, call
X-Rite
for
advice on how to replace it.
To perform a calibration, proceed as follows:
1. Select
CAL
from the ColorEye XTH
Main Menu
.
2. The following message appears:
PRIMARY CALIBRATION
Place White
Calibration Standard In
Position and Press
the ENTER key
3. Remove the white calibration standard from the pocket of the carrying strap. It is
tethered to the strap.
4. Close the target foot and latch it in place.
5. Position the white standard at the aperture opening. See Figure 4-1. Make sure
that the standard is properly aligned with the ridges on the aperture foot. Locate the
Top Alignment Notch on the standard and make sure it is aligned with the top of the
aperture foot. See Figure 4-2.
Figure 4-1. Calibration Standard Alignment
Top Alignment Notch
Aperture Foot
Alignment Notches
Aperture Foot
Alignment Notches
Calibration Standard
Instrument Showing Aperture Foot
Aperture Standard Alignment Ridge (4)