Page 10 of 18
52396 R00
FAULT INDICATIONS & TROUBLESHOOTING
HANDSET FAULT INDICATION
Beeping will be heard when pushing a button on the handset if there is a
problem with the handset itself or the associated control or motor system.
Contact Wy’East Medical for technical support should this occur.
JOTSTICK FAULT INDICATION
Flashing of the horizontal or vertical row of lights will appear if there is a
problem with the joystick itself or the associated control or motor system.
Contact Wy’East Medical for technical support should this occur.
TROUBLESHOOTING
Symptom
Possible Cause
Action
No power for driving or lifting
Battery is not fully charged
Charge battery
Battery is damaged
Replace battery
Red Emergency Stop switch is pushed in
Turn switch clockwise to reset
Unresponsive steering or jerky motion
when driving
One or both wheel locking levers not in
locked (down) position
Push both wheel locking levers
to locked (down) position
Battery will not charge
Battery is damaged
Replace battery
Light flashing on drive control joystick
Charger is connected to battery
Disconnect charger from battery
Sling bar will not raise or lower
Slack in sling bar strap
Remove slack in strap
Bad actuator or controls
Contact Wy’East for service
Sling bar will not raise
Twist in sling bar strap
Remove twist in strap