MAINTENANCE
PERIODIC MAINTENANCE
The following scheduled maintenance and inspections will help to ensure trouble-free operation of your Denali
600 Patient Lift.
Not recommended for use in shower or a wet environment.
Do not pressure-wash.
EVERY DAY
Charge the battery.
ONCE A MONTH
Check the sling bar strap for damage or fraying.
Check for loose or missing fasteners or other parts.
ONCE A YEAR
Perform load testing.
EVERY 2 YEARS
Replace the sling bar strap.
There are no components on the Denali 600 that require periodic lubrication. All bearings and moving
components are lubricated for the life of the product.
WARNING
No unauthorized modification of this equipment is allowed. Contact Wy’East Medical prior to any modification of
this equipment
CLEANING
Clean the unit once a month or as needed. For cleaning and disinfection use a 5% chlorine bleach dilution and an
EPA registered Hard Surface Disinfectant on painted and plastic surfaces. On all other surfaces use an EPA
registered Hard Surface disinfectant wipe that does not contain sodium hypochlorite (bleach).
The lift strap may only be cleaned with water using a damp cloth. Allow to dry completely before use.
Always test cleaning agents for compatibility on an inconspicuous area.