Operation
18
Aligning a marked
cell
A marked cell is aligned as follows:
Aligning an
unmarked cell
An unmarked cell is aligned as follows:
Balancing cells
Even insignificant variations in the glass can affect the mea-
sured value. For this reason, always use the same cell or
balance a pair of cells for low turbidity values:
1 Insert the cell in the cell shaft on the turbidity measur-
ing instrument
2 Align the marked cell so that the arrow on the marker
ring points to the marker pin on the casing of the cell
shaft.
1 Insert the cell in the cell shaft on the turbidity measur-
ing instrument
2 Slowly rotate the cell through one complete rotation
(by 360°)
3 Watch the display of the turbidity measuring instru-
ment while rotating the cell. Leave the cell in the po-
sition that gives the lowest display.
1 Fill the first cell with the sample, align it and mark it
(see 4.1.1 M
ARKING
AND
ALIGNING
CELLS
)
2 Press the
key to store the measured value
3 Fill the second cell with the same sample and insert
it in the cell shaft
4 Rotate the cell until the same measured value is dis-
played as for the first cell
5 Mark the cell with a marker ring.
Summary of Contents for Turb 550
Page 8: ...Overview 8...
Page 12: ...Safety 12...
Page 16: ...Commissioning 16...
Page 25: ...Operation 25 Note The flow through vessel can remain installed until the next calibration...
Page 42: ...Maintenance cleaning disposal 42...
Page 46: ...What to do if 46...
Page 50: ...Accessories options 50...
Page 54: ...Abbreviations and index 54...