Turb 2000 Series
Instrument configuration (CONFIG mode)
ba75695e03 10/2013
31
6.15 Desiccant
alarm
When the humidity detector in the Turb 2000 series indicates that the internal environment is close to
the point where humidity could cause condensation, the instrument will display DESC as a screen
warning.
If desired, a desiccant warning can:
•
Activate the alarms relays.
•
Can activate an alarm condition on the 4-20mA.
To activate the alarm relays when the desiccant fails, select set one or both alarms to Error (see section
6.5 C
ONFIGURING THE ALARMS
). To activate an alarm condition on the 4-20 mA set the ERLV to one of
the three alarm states (see section 6.3 C
ONFIGURING THE
E
RROR
L
EVEL
)
For either alarm modes to activate On must be selected in the DESC menu. The default for this menu is
OFF. Make selections using the
and
buttons then press the
OK
button to move to 4-20 mA
calibration.