Turb 2000 Series
Instrument configuration (CONFIG mode)
ba75695e03 10/2013
23
6
Instrument configuration (CONFIG mode)
The instrument has been designed to provide the ability to customize the instrument according to needs
at any time during normal operation. This mode has been split into submenus to facilitate instrument
configuration. This section describes how to use each of the submenus to configure the instrument.
While in the configuration mode, the instrument has a time-out feature that automatically returns the
system operation to the AUTO mode after a fifteen (15) minute period.
Select the CONFIG mode of the instrument by pressing the
MODE/EXIT
button until the arrow beside
CONFIG is illuminated on the display. Then press the
OK
button.
Note
To exit the CONFIG mode, press the MODE/EXIT button.
6.1
Selecting the output (O/P)
The first configuration selection is the O/P. The selections are 4-20 for the 4-20 mA output, 485 for the
RS-485 and OFF if no outputs are required. Select the desired output by using the
and
buttons.
Once the desired output has been set, press the
OK
button to accept it. The next prompts will depend
on the output selected.
Note
After one output has been selected the other output is rendered non-operable.
6.2
Setting the 4-20 mA output
To connect the output, see section 2.3.2. If the 4-20 mA output was turned on, prompts to set the 4 mA
(4MA) and 20 mA (20MA) turbidity limits levels will be displayed. There is also a menu to adjust the
error level (ERLV). The first prompt will be the turbidity limit assigned to the 4 mA output level: