Turb 2000 Series
Instrument configuration (CONFIG mode)
ba75695e03 10/2013
27
6.5.1 Alarm
1
Alarm 1 Function: The ALM1 is displayed and the display indicates the current function of alarm 1 (HI,
LO, OFF or Error). Use the
or
buttons to cycle through and select the desired function. Press the
OK
button to accept the selection.
If the alarm was turned OFF a prompt will appear to set up alarm 2 (go to section 6.5.2). If, on the other
hand, one of the other functionalities was selected a prompt will appear to set the delay times.
Alarm 1 set point: This prompt is used to select the set point for this alarm; this is indicated by “S/P”
shown on the lower row of the display. Select the desired alarm level by using the
and
buttons.
Once the desired output has been set, press the
OK
button to accept it.
Alarm 1 delay times:
Delay on
: The following display will appear to allow to select the number of
seconds currently set for the “delay on” time.
The current selected number of seconds will be shown. Select the desired number of seconds for the
“delay on” time for this alarm using the
and
buttons. Once the desired delay time has been set,
press the
OK
button to accept it.
Delay off
: Next, the following display will appear to select the number of seconds currently set for the
“delay off” time.
The current selected number of seconds will be shown. Select the desired delay off time for this alarm
using the
and
buttons. Once the desired delay time has been set, press the
OK
button to accept
it. After the settings for alarm 1 have been completed, prompts will allow for the set up of the
information on Alarm 2.
6.5.2 Alarm
2
Repeat the procedure listed in section 6.5.1 to set up the parameters for alarm 2. If one of the other
functionalities is selected, a prompt to set the delay times and the set point, as with Alarm 1, will be
displayed.