Error elimination
Turb 2000 Series
36
ba75695e03 10/2013
8.3 Error
messages
Symptom Cause
Remedy
Lower display shows CAL
Instrument cannot be
calibrated
See sections 4.3 and 5.2
Lower display shows MA
4-20 mA loop open
Check wiring. See sections
2.3.2 and 6.2
Lower display shows CLN
No contact to ultrasonic
cuvette
See section 7.1
Lower display shows DESC
Desiccant pouch bad
Change desiccant pouch. See
section 9.2
Lower display shows LAMP
Lamp failed
The lamp has to be replaced.
Refer to section 9.3
Lower display shows FLOW
Sample flow has stopped
Restore flow. Contact WTW
about factory installed option
Lower display shows FAIL
System error
Refer to section 8.1 & 8.2
Readings are higher than
expected
Bubbles in solution
Ensure that the drain vent is
open and is not obstructed. See
section 2.2.2
Apply backpressure. See
section 2.2 and figure 4
For severe cases of bubbles a
stilling chamber is available. Call
WTW.
Condensate or leaky cuvette
Check flow through cuvette for
condensate or leaks.
Flow through cuvette dirty
Clean the cell. See section 9.1
Instrument out of calibration
Recalibrate. Refer to section 4
Readings are erratic
Bubbles in solution
See above
Debris in flow through
Clean debris from cuvette
Readings are lower than expected Instrument out of calibration
Recalibrate. Refer to section 4
Upper display flashes
Sample over-range
Check sample. Sample may be
too high to read.